Printer setup, Automatic prints, Printing waveforms – ZOLL Propaq MD Rev A User Manual

Page 190

Advertising
background image

C

HAPTER

17

P

RINTING

17–2

www.zoll.com

9650-0805-01 Rev. A

Printer Setup

To configure the printer functions, press the More quick access key (

), press the Setup quick

access key (

), highlight Printer, and press Select. Use the navigation keys to select the

desired print configuration. The Setup>Printer window contains the following print
configuration options:

Automatic Prints

You can print any following events by setting the monitor to print automatically in the
Setup>Supervisor>Printer window. To do this, press (

), then press

. Using the

navigation buttons, select Supervisor, and then Printer.

Defib/Pacer Events

Event

Treatment Snapshot

Treatment Summary Snapshots

Printing Waveforms

You can print waveforms by pressing the Print quick access key (

). The Propaq MD unit

will print the displayed waveforms as configured in the Number of Traces option. The Print
Number of Traces option can be set to

1

,

2

, or

3

. Patient vital sign numerics are printed above

the waveforms.

Note:

Waveforms that have invalid data print a dashed line for that trace.

The ECG Waveform

ECG waveforms are printed on a grid with major divisions (dotted lines) every 5mm and minor
divisions (single dots) every 1 mm. The ECG waveform is always printed if ECG is monitored.

The Invasive Pressure Waveform

Pressure waveforms are printed on a grid with major divisions (vertical dotted lines) every 5
mm, and the pressure scale grids are printed horizontally.

Print configuration

Options

Print Speed

25 mm or 50 mm per second

Print Number of Traces

1, 2, or 3

Print Grid

On or Off

Print on Snapshot

On or Off

Print on Patient Alarm

On or Off

Print on NIBP

On or Off

Print on Trend

On or Off

Advertising