1 select energy level, 2 charge defibrillator – ZOLL Propaq MD Rev A User Manual

Page 174

Advertising
background image

M

ANUAL

D

EFIBRILLATION

14-12

www.zoll.com

9650-0805-01 Rev. A

Note:

The unit can be configured to stay in SYNC mode after defibrillation in
Defib/Pacer default settings in the Setup>Supervisor menu.

1 Select Energy Level

Press the Select Energy arrows up or down to select the desired energy level.These buttons are
located either on the front of the unit or on the STERNUM paddle.

Warning!

When used with pedi-padz, defibrillator energies must be set manually based on site-
specific institutional protocols for pediatric defibrillation.

2 Charge Defibrillator

Press the CHARGE button on the front panel or on the APEX paddle handle.

To increase or decrease the selected energy after you have pressed the Charge button, use the
defibrillator Select Energy arrows on the front panel or sternum panel.

Caution

Changing the selected energy while the unit is charging or charged causes the defibrillator to
disarm itself. Press the CHARGE button again to charge the unit to the newly selected energy
level.

A charging message displays at the bottom of the screen, and a distinctive charging tone sounds
indicating that the unit is charging. The energy range bar graph on the right side of the display
highlights the charging progress until it reaches the selected energy. When the unit is fully
charged, the tone changes to a continuous charge ready tone, the highlighted energy bar graph
includes the selected energy, and the shock button lights up.