IM 01E20F02-01E

5-12

5. EXPLANATION OF BASIC ITEMS

10:uP Fault

11:EEPROM Fault

12:A/D(H) Fault

13:A/D(L) Fault

14:A/D(Z) Fault

15:Coil Open

16:EEPROM Dflt

100:Comm uP Fault

101:Comm EEPROM Fault

102:IT1 Not Saved

103:IT2 Not Saved

104:Comm Error1

105:Comm Error2

106:DL Incomplete

107:Download Fail

108:Not Ready

30:Sig Overflow

31:Empty Pipe

33:Adhesion Alm

110:AI Lo Lo Alm

111:AI Hi Hi Alm

112:PID Lo Lo Alm

113:PID Hi Hi Alm

40:RS O/S Mode

41:TB O/S Mode

42:AI FB O/S Mode

43:IT1 FB O/S Mode

44:IT2 FB O/S Mode

45:DI1 FB O/S Mode

46:DI2 FB O/S Mode

47:AR FB O/S Mode

48:PID FB O/S Mode

50:Span > 10m/s

51:Span < 0.1m/s

72:Size Set Err

120:IT1 Clock Per Err

121:IT2 Clock Per Err

122:AR Range Set Err

Microprocessor (CPU) failure

EEPROM failure

A/D converter failure

Flowtube coil is open-circuit

EEPROM default values

Communication uP failure

Communication EEPROM failure

IT1 save error

IT2 save error

AXF internal communication error

AXF internal communication error

Download is not completed

Download failure

Function block not scheduled

Input signal error

Flowtube is not filled with fluid

Electrode adhesion alarm

AI process alarm

AI process alarm

PID process alarm

PID process alarm

RS. MODE_BLK.Target is O/S mode

TB. MODE_BLK.Target is O/S mode

AI. MODE_BLK.Target is O/S mode

IT1. MODE_BLK.Target is O/S mode

IT2. MODE_BLK.Target is O/S mode

DI1. MODE_BLK.Target is O/S mode

DI2. MODE_BLK.Target is O/S mode

AR. MODE_BLK.Target is O/S mode

PID. MODE_BLK.Target is O/S mode

Span flow velocity setting is 11 m/s or more

Span flow velocity setting is 0.05 m/s or less

Mass units have been selected for Base

Flow Unit but density is set to zero.

Measure Mode is set to Enhanced DF without

selecting an optional code HF1 or HF2.

A value of 3000.1 mm or more is set for Nominal Size.

The condition in Adhesion detection level,

Level:1<Level:2<Level:3<Level:4 is not satisfied.

IT1 CLOCK_PER set value is smaller than excecution period

IT2 CLOCK_PER set value is smaller than excecution period

RANGE_HI>RANGE_LO is not satisfied

Check XD Scale,Density Unit,Mass Flow Density

Check Measure Mode of TB

Check Nominal Size,Nominal Size Unit

Check Adhesion Level1 to Adhesion Level 4

Check Clock Period,Period of Execution

Check Clock Period,Period of Execution

Check Range Hi and Range Lo of AR

Check XD Scale of AI

Schedule FB, or check LAS communication

Check signal cable and grounding

Fill flow tube with fluid

Clean electrodes

Check the flow rate and setting value

Check the flow rate and setting value

Check the setting value

Check the setting value

Check software download error code

Cut the power and check coil & EX cables

Contact nearest office or service center

Contact nearest office or service center

Category

Category

Alarm Message

Alarm Message

Countermeasure Message

Countermeasure Message

Alarm Description

Alarm Description

Process

Alarms

O/S Mode

Alarms

Setting
Alarms

System

Alarms

T0515.eps

73:Adh Set Err

71:Meas Mod Set

57:Dens Set Err

Warning Display