APPENDIX 2. INTEGRATOR (IT) BLOCK

APPENDIX 2. INTEGRATOR (IT) BLOCK

IN_1: Block input 1 (value and status)

IN_2: Block input 2 (value and status)

REV_FLOW1: Indicates whether the sign of IN_1 is
reversed. It is a discrete signal.

REV_FLOW2: Indicates whether the sign of IN_2 is
reversed. It is a discrete signal.

RESET_IN: Resets the integrated values. It is a
discrete signal.

RESET_CONFIRM: Reset confirmation input. It is a
discrete signal.

OUT: Block output (value and status)

OUT_PTRIP: Set if the target value exceeds
PRE_TRIP. It is a discrete signal.

OUT_TRIP: Set if the target value exceeds
TOTAL_SP (or 0). It is a discrete signal.

The Integrator (IT) block adds two main inputs and
integrates them for output. The block compares the
integrated or accumulated value to TOTAL_SP and
PRE_TRIP and generates discrete output signals
OUT_TRIP or OUT_PTRIP when the limits are
reached.

The output is as represented by the following equation
(for counting upward and rate conversion).

OUT.Value = Integration start value + Total
Total = Total + Current Integral
Current Integral = (x + y)

ϫ ∆t

x: IN_1 value whose unit has been converted
y: IN_2 value whose unit has been converted
∆t: block execution period

A2.1 Schematic Diagram of Integrator Block

The following shows the schematic diagram of the Integrator block.

OUT

IN_1

RESET_IN

REV_FLOW1

–1

TIME_UNIT2

IN_2

REV_FLOW2

PULSE_VAL2

–1

UNIT_CONV

OP_CMD_INT

(RESET)

INTEG_OPTS

(CARRY)

Reverse

Forward

Forward

OUT_PTRIP

OUT_TRIP

INTEG_OPTS
(FLOW TYPE)

Add

Compare

Compare

TOTAL / RTOTAL

Integrate

INTEG_TYPE

INTEG_OPTS (QUALITY)

GOOD_LIM

UNCERT_LIM

CLOCK_PER

N_RESET

MAN

MAN

MAN

Convert Rate

Convert Accum

Convert Rate

Convert Accum

TOTAL_SP

PRE_TRIP

INTEG_OPTS
(INPUT TYPE)

PULSE_VAL1

TIME_UNIT1

INTEG_OPTS
(INPUT TYPE)

RESET_CONFIRM

Reverse

FA0201.EPS

Figure A2.1 Integrator Block

The Integrator block is classified into the following
five sections for each function:

• Input process section: Determines the input value

status, converts the rate and accumulation,
and determines the input flow direction.

• Adder: Adds the two inputs.

• Integrator: Integrates the result of the adder into the

integrated value.

• Output process section: Determines the status and

value of each output parameter.

• Reset process section: Resets the integrated values.