8. EXPLOSION PROTECTED TYPE INSTRUMENT

8.4 IECEx

WARNING

Only trained persons use this instrument in
industrial locations.

(1) Technical Data

*AXF002C – AXF400C

Applicable Standard:

IEC60079-0: 2004, IEC60079-1: 2003,
IEC60079-7: 2001, IEC60079-11: 1999,
IEC60079-18: 2004,
IEC61241-0: 2004, IEC61241-1: 2004,
IEC60529: 1999 + Edition 2.1: 2001

Certificate: IECEx KEM 05.0018

(Integral Flowmeter)
IECEx Flameproof Type

Ex demb[ia] IIC T6...T3
Electrode Circuit Um: 250 Vac/dc
Maximum power supply voltage: 250 Vac/130 Vdc
Excitation Circuit: 140V max
Enclosure: IP66, IP67
Temperature Class:

T0809.EPS

Temperature

Class

–40

°C to +85°C (–40°F to +185°F)

–40

°C to +130°C (–40°F to +266°F)

–40

°C to +120°C (–40°F to +248°F)

–40

°C to +70°C (–40°F to +158°F)

T6

Process Temperature

T4

T5

T3

Ambient Temp.:

PFA Lining; –40˚C to +60˚C (–40˚F to +140˚F)
Ceramics Lining; –15˚C to +60˚C (+5˚F to +140˚F)

IECEx Type of Protection "Dust"

Ex tD A21 IP6x T95˚C, T105˚C, T120˚C, T130˚C
Electrode Circuit Um: 250 Vac/dc
Maximum power supply voltage: 250 Vac/130 Vdc
Excitation Circuit: 140V max
Enclosure: IP66, IP67
Maximum surface temperature:

T0810.EPS

Maximum Surface

Temperature

–40

°C to +85°C (–40°F to +185°F)

–40

°C to +130°C (–40°F to +266°F)

–40

°C to +120°C (–40°F to +248°F)

–40

°C to +70°C (–40°F to +158°F)

T95

°C (+203°F)

Process Temperature

T120

°C (+248°F)

T105

°C (+221°F)

T130

°C (+266°F)

Ambient Temp.:

PFA Lining; –40˚C to +60˚C (–40˚F to +140˚F)
Ceramics Lining; –15˚C to +60˚C (+5˚F to +140˚F)

(Remote Flowtube)
IECEx Flameproof Type

Ex demb[ia] IIC T6...T3
Electrode Circuit Um: 250 Vac/dc
Excitation Circuit: 170V max
Enclosure: IP66, IP67
Temperature Class:

T0811.EPS

Temperature

Class

–40

°C to +85°C (–40°F to +185°F)

–40

°C to +150°C (–40°F to +302°F)

–40

°C to +120°C (–40°F to +248°F)

–40

°C to +70°C (–40°F to +158°F)

T6

Process Temperature

T4

T5

T3

Ambient Temp.:

PFA Lining; –40˚C to +60˚C (–40˚F to +140˚F)
Ceramics Lining; –15˚C to +60˚C (+5˚F to +140˚F)

IECEx Type of Protection "Dust"

Ex tD A21 IP6x T95˚C, T105˚C, T120˚C, T135˚C
Electrode Circuit Um: 250 Vac/dc
Excitation Circuit: 170V max
Enclosure: IP66, IP67
Maximum surface temperature:

T0812.EPS

Maximum Surface

Temperature

–40

°C to +85°C (–40°F to +185°F)

–40

°C to +150°C (–40°F to +302°F)

–40

°C to +120°C (–40°F to +248°F)

–40

°C to +70°C (–40°F to +158°F)

T95

°C (+203°F)

Process Temperature

T120

°C (+248°F)

T105

°C (+221°F)

T135

°C (+275°F)

Ambient Temp.:

PFA Lining; –40˚C to +60˚C (–40˚F to +140˚F)
Ceramics Lining; –15˚C to +60˚C (+5˚F to +140˚F)

(2) Installation

WARNING

• All wiring shall comply with local installation

requirements and local electrical code.

• In hazardous locations, the cable entry devices

shall be of a certified IECEx flameproof type,
suitable for the conditions of use and correctly
installed.

• Unused apertures shall be closed with suitable

flameproof certified blanking elements. (The
plug attached is certified as the flameproof and
IP66 or IP67 as a part of this apparatus.)

• In case of ANSI 1/2 NPT plug, ANSI hexagonal

wrench should be applied to screw in.