6 sanitary type, Sanitary type -22 – Yokogawa ADMAG AXF User Manual

Page 32

Advertising
background image

IM 01E20D01-01E

3-22

3. INSTALLATION

3.3.6

Sanitary Type

The sanitary type can be mounted to the piping using a
clamp, a union, or a welded joint.

NOTE

This section describes the remote flowtube as an
example. The same procedure also applies to
the integral flowmeter.

Sanitary adapters are engraved with character as
indentification.

Table 3.3.20

Sanitary Adapter Indentification

T3.3.20.EPS

No mark

F

L

FL

SUS304

SUSF304

SUS316L

SUSF316L

Material for Adapter

Indentification Character

(1) Mounting Direction

Mount the flowmeter so that the flow direction of the
fluid to be measured is in line with the direction of the
arrow mark on the flowmeter.

IMPORTANT

If it is impossible to match the direction of the
arrow mark, the direction of the electrical con-
nection can be changed. Refer to Section 5.1 to
do this properly.
In case the fluid being measured flows against
the arrow direction, refer to the parameter J20:
Flow Direction
in the user’s manual of the
AXFA11 Magnetic Flowmeter Remote Converter
(IM 01E20C01-01E) or the AXFA14 Magnetic
Flowmeter Remote Converter/AXF Integral
Flowmeter [Software Edition] (IM 01E20C02-
01E).

(2) Mounting Procedure

(a) Clamp type

(process connection codes: HAB, HDB, and HKB)

1) Welding ferrule

Weld a ferrule to the piping.

IMPORTANT

When welding the ferrule, pay attention to the
edge preparation, level differences between the
ferrule and the piping, and the welding current to
avoid deforming the piping or causing stagnation
portion of the fluid.

2) Installing gasket

Install a gasket for clamp to fit in the groove on
the ferrule.

3) Positioning flowmeter

Position the flowmeter between the two ferrules.

4) Tightening clamp

Install a clamp to cover the tapered parts of the
flowmeter-side adapter and the ferrule, and
tighten the clamp screw.

5) Confirmation of adapter mounting screw

After installation of the magnetic flowmeter, be
sure to retighten the adapter mounting screw
according to Table 3.3.21 or Table 3.3.22.
Be sure to confirm that leakage from adapter
connection point does not occur by filling the
pipe of the flowtube with fluid.

NOTE

The ferrule, clamp, and gasket are not provided
with the flowmeter, and must be provided by the
user.

Gasket

Ferrule

Clamp

Adapter for clamp connection

Adapter mounting screw

F0314.EPS

Figure 3.3.8

Mounting Procedure for Clamp Connection
Type