4. WIRING

4.2.2

Cables

(1) Dedicated Signal Cable (AXFC)

Outer jacket

10.5 (0.413")

Tape

Shield (C)

Insulation

Insulation

Shields (SA and SB)

Conductors (A and B)

F0417.EPS

Figure 4.2.1

Dedicated Signal Cable AXFC

The flow signal is transmitted via this dedicated cable.
The cable is constructed with double shielding over the
two conductors, and heat-resistant vinyl is used for the
outer jacket material.

Finished diameter:

10.5 mm (0.413")

Maximum length:

Combination with the AXFA11 converter:

200 m (660 ft)

Combination with the AXFA14 converter:

100 m (330 ft)

Maximum temperature:

80

°C (176°F)

IMPORTANT

If the cable is longer than required, cut off any
extra length rather than coiling it up, and termi-
nate the conductors as shown in Figure 4.2.2.
Avoid using junction terminal boards to extend
the cable length, as this will interrupt the shield-
ing.

20 (0.8)

10.5 (0.4)

AXFC

L (Specified Dimensions)

150

Ϯ

5

8 (0.3) max.

150

Ϯ

5

70 (2.76)

60 (2.36)

90 (3.54)

(5.9)

(5.9)

55 (2.17)

90 (3.54)

50 (1.97)

25 (0.98)

SA

SB

A

C

B

A

C

B

8(0.3) max.

White Black Red

White Black

Red

F0418.EPS

Unit : mm

(approx. inch)

On the
converter
side

On the
flowtube
side

Figure 4.2.2

Treatment of Dedicated Signal Cables

CAUTION

• As crimp terminals A, B, SA, SB and C have

their own electrical potentials, securely insulate
them so as not to come in contact with one
another.

• To prevent a shield from coming in contact with

another shield or the case, cover each shield
with a vinyl tube or wrap it in vinyl tape.

NOTE

Conductors A and B carry the signal from the
electrodes, and C is at the potential of the liquid
(signal common). Shields SA and SB are kept
at the same potentials as the individual elec-
trodes (these are actively driven shields). This is
done to reduce the effect of the distributed
capacitance of the cable at long cable length.
Note that, since the signals from the individual
electrodes are impedance converted inside the
converter, errors will result if they come in
contact with any other component. Great care
must be taken in the cable end treatment.

(2) Excitation Cable

JIS C3401 control cable equivalent
JIS C3312 power cable equivalent
14 AWG Belder 8720 equivalent

Outer Diameter:

With no gland option:

6.5 to 12 mm (0.26 to 0.47 in.)

With gland options EG, EU and EW:

10.5 or 11.5 mm (0.41 to 0.45 in.)

With gland options EP:

6 to 12 mm (0.24 to 0.47 in.)

Nominal Cross Section:

Single wire; 0.5 to 2.5 mm

2

Stranded wire; 0.5 to 1.5 mm

2