4 wiring connections, Wiring connections -11 – Yokogawa ADMAG AXF User Manual

Page 45

Advertising
background image

IM 01E20D01-01E

4-11

4. WIRING

F0421.EPS

Gasket

Washer

Tightening gland

When working on conduit pipes or flexible pipes (PF1/2 only)

Cable

PF1/2

Figure 4.2.5

Waterproof Gland with Union Joint
(Optional code EU)

Plastic gland

F0422.EPS

Gasket

Figure 4.2.6

Plastic Gland (Optional code EP)

(3) Conduit Wiring

When wiring the conduits, pass the conduit through the
wiring connection port, and utilize the waterproof
gland to prevent water from flowing in. Place the
conduit pipe on an angle as shown in Figure 4.2.7.
Install a drain valve at the low end of the vertical pipe,
and open the valve regularly.

F0423.EPS

Drain valve

Figure 4.2.7

Conduit Wiring

4.2.4

Wiring Connections

WARNING

Before wiring, be sure that the power supply for
AXFA11 or AXFA14 converter has been turned
off to prevent an electrical shock.

(1) Removing Cover (Size 2.5 to 1000 mm (0.1

to 40 in.))

Loosen the cover locking screw clockwise using a
hexagonal wrench (nominal size 3) to unlock the cover.
(Upon shipment from the manufacturing plant, the
cover is unlocked.) Hold the flowtube with your hand
and remove the cover by turning it in the direction of
the arrow as shown below.

F0424.EPS

Figure 4.2.8

Removing the Terminal Box Cover (Remote
Flowtube)

(2) Terminal Configuration

When the cover is removed, the connection terminals
will be visible.

Terminal

Symbols

A
B
C

EX1
EX2

Flow signal output

Excitation current input

Protective grounding
(Outside of the terminal)

Description

F0425.EPS

Figure 4.2.9

Terminal Configuration (General-Purpose
Use, Submersible Type, Sanitary Type, Size
2.5 to 400 mm (0.1 to 16 in.))

Advertising