IM 01E20D01-01E

3-6

3. INSTALLATION

*: To be provided by user

F0309a.EPS

*Mounting base

Gasket

Connecting fitting

Union joint nut

*Piping

Horizontal mounting

F0309b.EPS

Vertical mounting

F0309c.EPS

Figure 3.3.1

Mounting Procedure for Union Joint Type

Table 3.3.2

Tightening torque values for Union Joint
Type

Apply these tightening torque values when the gaskets
are Valqua #7020 (standard) or alkali-resistant gaskets
for the metal piping (optional code GF).

T0303.EPS

Torque (N-m / {kgf-cm} / [in-lbf])

Size mm (inch)

9 to 12 / {91.77 to 122.4} / [79.66 to 106.2]

9 to 12 / {91.77 to 122.4} / [79.66 to 106.2]

14 to 18 / {142.8 to 183.5} / [123.9 to 159.3]

2.5 (0.1)

10 (0.4)

5 (0.2)

Table 3.3.3

Tightening torque values for Union Joint
Type and Permeable Fluids

T0304.EPS

Torque (N-m / {kgf-cm} / [in-lbf])

Size mm (inch)

11 to 15 / {112.2 to 153} / [97.36 to 132.8]

11 to 15 / {112.2 to 153} / [97.36 to 132.8]

17 to 23 / {173.4 to 234.5} / [150.5 to 203.6]

2.5 (0.1)

10 (0.4)

5 (0.2)

3.3.2

Nominal Diameter 2.5 mm (0.1
in.) to 40 mm (1.5 in.), Wafer
Type

IMPORTANT

Use bolts and nuts in compliance with the flange
ratings. When stud-type through-bolts are used,
be sure the outside diameter of the shank is
smaller than that of the thread ridge. Be sure to
choose a gasket with inner and outer diameters
that does not protrude inside the piping (refer to
Table 3.3.13). If the inner diameter of the gasket
is too large, or outer diameter of the gasket is
too small, fluid leakage may result.

(1) Mounting Direction

Mount the flowmeter so that the flow direction of the
fluid to be measured is in line with the direction of the
arrow mark on the flowmeter.

IMPORTANT

If it is impossible to match the direction of the
arrow mark, the direction of the electrical con-
nection can be changed. Refer to Section 5.1 to
do this properly.
In case the fluid being measured flows against
the arrow direction, refer to the parameter J20:
Flow Direction
in the user’s manual of the
AXFA11 Magnetic Flowmeter Remote Converter
(IM 01E20C01-01E) or the AXFA14 Magnetic
Flowmeter Remote Converter/AXF Integral
Flowmeter [Software Edition] (IM 01E20C02-
01E).

(2) Mounting Centering Devices

To maintain concentricity of the flowmeter with the
pipes, install centering devices on the Mini-flanges of
the flowmeter. Use the appropriate centering devices
according to the nominal diameter and the flange
ratings.