Hazardous area classification, Standard performance, Accuracy – Yokogawa ADMAG AXF User Manual
Page 71

IM 01E20D01-01E
6-9
6. OUTLINE
Natural Soft Rubber, EPDM Rubber lining:
Internal insertion type
Ceramics lining: Integral type
Sanitary Type: Internal insertion type
Replaceable Electrode Type
Electrode parts can be put into unit to facilitate replace-
ment or mounting at customer site.
The optional dedicated tool (F9807SK) is required.
Replaceable electrodes are available for the
following:
AXF standard:
Use
Lining
Wafer
Flange
25 to 300 mm
(1.0 to 12 in.)
25 to 400 mm
(1.0 to 16 in.)
General-
Purpose Use
Process
Connection
Available
Size
Electrode
Material
PFA/
Polyurethane
Rubber
JIS SUS316L
(AISI 316L
SS/EN
1.4404
equivalent)
(
*
1)
T06.EPS
Replacement model for earlier ADMAG or ADMAG AE:
Use
Lining
Flange
150 to 250 mm
(6.0 to 10 in.)
Process
Connection
Available
Size
Electrode
Material
PFA/
Polyurethane
Rubber
JIS SUS316L
(AISI 316L
SS/EN
1.4404
equivalent)
(
*
1)
T07.EPS
General-
Purpose Use
*1: If any other electrode materials are required, please
contact Yokogawa office.
Mounting/Shapes (Remote Flowtube):
• Electrical Connection: ANSI 1/2 NPT female
ISO M20
ϫ 1.5 female
JIS G1/2 (PF1/2) female
• Direction of Electrical Connection: The direction can be
changed even after delivery.
Note: In case of submersible types, an optional code DHC,
and sizes of 1100 mm or larger, the direction can not
be changed after delivery.
• Terminal Connection at Terminal Box: M4 size screw
Grounding:
Grounding resistance 100
Ω or less
* In case of explosion proof type except TIIS, follow the
domestic electrical requirements as regulated in each
country.
* In case of TIIS Flameproof type, refer to description of
“HAZARDOUS AREA CLASSIFCATION”.
Combined Converter:
• A remote flowtube for sizes of up to 400 mm can be
combined with the AXFA11 Converter or the AXFA14
Converter. If a combined converter is changed from
AXFA11 to AXFA14 or vice versa, a new meter factor
must be adjusted by flow calibrations.
• In case that size 250 mm (10 in.) or larger is used in low
conductivity or high concentration slurries, please use the
AXFA11 Converter.
• A remote flowtube for sizes of 500 mm or larger can be
combined with the AXFA11 Converter only.
• Maximum Cable Length:
Combination of AXF remote Flowtube and AXFA11:
up to 200 m (660 ft)
Combination of AXF remote Flowtube and AXFA14:
up to 100 m (330 ft)
■ HAZARDOUS AREA CLASSIFICATION
Refer to Chapter 8.
■ STANDARD PERFORMANCE
Accuracy
Note: The accuracy of a product before shipment is
defined as totalized value at the result of calibration
test in our water actual flow test facility.
Calibrated conditions in our water actual test facility
are as follows:
Fluid temperature;
20
± 10°C
Ambient temperature; 20
± 5°C
Length of straight runs;10 D or more on the
upstream side; 5 D or more
on the downstream side
Reference conditions; Similar to BS EN29104
(1993); ISO 9104 (1991)
Pulse Output:
PFA/Ceramics Lining;
Size mm
(in.)
V
Ͻ 0.3 (1)
Ϯ1.0 mm/s
Ϯ0.5 mm/s
Ϯ0.5 mm/s
0.3
Ϲ V Ϲ 10
(1) (33)
V Ͻ 0.15 (0.5)
V Ͻ 0.15 (0.5)
0.15
Ϲ V Ϲ 10
(0.5) (33)
0.15
Ϲ V Ϲ 10
(0.5) (33)
V Ͻ0.15 (0.5)
0.15
Ϲ V Ͻ1
(0.5) (3.3)
1
Ϲ V Ϲ10
(3.3) (33)
Ϯ0.35% of
Rate
Ϯ0.5 mm/s
Ϯ0.35% of
Rate
Ϯ0.35% of
Rate
2.5 (0.1)
to
15 (0.5)
250 (10)
to
400 (16)
25 (1.0)
to
200 (8.0)
—
—
Flow Velocity
V m/s (ft/s)
Flow Velocity
V m/s (ft/s)
Standard
Accuracy
(Calibration
code B)
High Grade
Accuracy
(Calibration
code C)
Ϯ0.18% of Rate
Ϯ 0.2mm/s
Ϯ0.2% of Rate
T08.EPS
Enhanced dual frequency excitation(Option code HF2):
Standard accuracy
Ϯ
1 mm/ s
Size 2.5 mm (0.1 in.) to 15 mm (0.5 in.)
1.2
1.0
0.8
0.6
0.4
0.35
0.2
0
0
1
2
6
8
10
V[m/s]
High grade Accuracy
High grade Accuracy
Standard Accuracy
F14.EPS
Pulse
Output
Accuracy
% of Rate
Size 25 mm (1.0 in.) to 400 mm (16 in.)
1.2
1.0
0.8
0.6
0.4
0.2
0
0.35
0
1
2
6
8
10
V[m/s]
F15.EPS
Pulse
Output
Accuracy
% of Rate
Standard Accuracy
High grade Accuracy