IM 01E20D01-01E

5-4

5. MAINTENANCE

5.2.3

Installing Replaceable Elec-
trodes

NOTE

It is recommended to replace the O-ring when
reinstalling the replaceable electrode. Use the O-
ring specified by Yokogawa (G9303SE: material
is fluororubber).

NOTE

Precautions for storage of O-rings:
• Keep them in a cool, dark place.
• Wrap them well.
• Do not use O-rings after one year since their

purchase.

(1) Push the electrode holder with end B of the tool

screwed into it, straight into the flowtube.

F0510.EPS

Electrode holder:
(Electrode A is engraved with "A"
and electrode B has no markings.)

Electrode lead wire:

Red: Electrode A
White: Electrode B

NOTE

• Check that the O-ring has not come off.
• Do not interchange electrodes A and B when

installing them. Install electrode A to the
electrode boss with the red lead wire and
electrode B to the one with the white lead wire.

(2) Using end A of the tool, tighten the electrode

holder.

F0511.EPS

Torque driver

NOTE

Using the torque wrench or torque driver, tighten
the electrode holder to the following torque
values:

PFA lining:

8.0 N·m

±1 N·m (80 kgf·cm ±10 kgf·cm)

Polyurethane rubber lining:

3.0 N·m

±0.5 N·m (30 kgf·cm ±5 kgf·cm)

WARNING

If the electrode holder was not tightened to its
specified torque value, fluid leakage from the
electrode may result. Should fluid spill over the
electrode block, wipe it dry and check that the
O-ring has not come off.

(3) Fill the interior of the flowtube with the fluid and

check that there is no leakage from the electrode.

(4) Hold down the lead wire and terminal lug and

screw the mounting screw into the electrode using
the Phillips screwdriver.

F0512.EPS

Electrode mounting screw