4. WIRING

(4) Wiring the Remote Flowtube (Explosion

Proof Type) with Converters

IMPORTANT

In case of TIIS Flameproof type, a remote
flowtube is available for combined use with
AXFA14 converter only.

1) Connection with the AXFA11 converter

In case of explosion proof type for ATEX, FM, CSA
and IECEx certification, connect wiring as shown in
the figure below.

In case of the explosion proof type, the protective
grounding

of remote flowtube must be connected to

a suitable IS grounding system. In that case,
(functional grounding terminal) need not be connected.

I+

I–

CURRENT OUT

AL+

AL–

C

SA

A

B

SB

ALARM OUT

N/–

L/+

POWER SUPPLY

EX2

EX1

EXCIT ATION

P–

SI1+

SI2+

COM

P+

PULSE OUT

STATUS IN

SIGNAL

SO1+

COM

SO2+

STATUS OUT

FUSE

2.5A 250V

AXFA11 converter

AXFC dedicated
signal cable

Converter

Remote

flowtube

F0428.EPS

SA

A

B

SB

C

EX1

EX2

Taping*

A

B

Taping*

C

EX1

EX2

* Individually tape and insulate the

shields corresponding to SA and
SB on the remote flowtube side.

Excitation cable

EX2

EX1

A

B

C

Remote flowtube

Figure 4.2.15 Wiring Diagram

2) Connection with the AXFA14 converter

In case of explosion proof type for ATEX, FM, CSA,
IECEx and TIIS certification, connect wiring as shown
in the figure below.

In case of the explosion proof type, the protective
grounding

of remote flowtube must be connected to

a suitable IS grounding system. In that case,
(functional grounding terminal) need not be connected.

AXFC dedicated
signal cable

Excitation cable

EX1

EX2

C

SA

A

B

SB

F0429.EPS

Converter

Remote

flowtube

SA

A

B

SB

C

EX1

EX2

Taping*

A

B

Taping*

C

EX1

EX2

* Individually tape and insulate the

shields corresponding to SA and
SB on the remote flowtube side.

AXFA14 converter

EX2

EX1

A

B

C

Remote flowtube

Figure 4.2.16 Wiring Diagram