Item

Description

Page

HART

Communicator

Select the output direction/value upon Sensor failure from
Low, High, off or user-setting value (mA or %).

Show the current setting of the output direction in CPU failure
which is set by hardware switch on a CPU assembly.

Select a set of data to be continuously sent; (1)PV, (2)output
in % range & current, (3)PV and output in current.

Enable/disable the burst mode.

Setting the polling address (0 to 15).

Select variables/information to be displayed on an integral
indicator.

Select the update period form fast, normal, and slow.

Process variables.

% Output variable

4 to 20 mA Output variable

Terminal temperature variable

Used for loop checks. Output can be set from 4, 20 mA or the
designated value within 3.6 to 21.6 mA.

Check the transmitter status. If an error is detected, the
corresponding message is displayed.

Reset the CPU of the transmitter and check the status.

Display of the result of self-test.

Show/not show warning messages.

Up to 4 error histories are stored in EEPROM.

Maximum and Minimum value of PV, SV, TV, 4V and Terminal
Temperature during the operation.

Operation time from last power-up.

Unintentional power-loss during operation may be detected.

Displays the permit/protect status for setting changes.

Write protect status is released for 10 minutes when the
password is entered.

Setting a new password.

Shows if the “joker” password has been used during the
operation.

Select input trimming mode. The mode depends on the kind
of calibration device and the number of calibration points.

Select the action type of input trimming.

Input an additional compensation to the factory set
characterization curve stored in the transmitter.

Adjust the 4 mA to 20 mA output.

Yokogawa
YTA110/YTA310/YTA320

Lower Limit for PV.
Upper Limit for PV.
Lower Limit for Sensor 1(2) input.
Upper Limit for Sensor 1(2) input.
serial number information
Lower Limit for terminal temperature.
Upper Limit for terminal temperature.

Snsr burnout type

AO Alrm typ

Burst option

Burst mode

Poll addr

Process Disp,
%/mA Disp,
Err-No Disp,
Bar graph,
Matrix Disp

Disp update

PV,SV,TV,4V

PV % rnge

PV AO

Term

Loop test

Self test

Master test

Status

Set warning enbl

Error log view

Max/Min log

Operate Time

Power Check

Write protect

Enable wrt
10 min

New password

Software Seal

Input trimming
mode

Snsr1(2) trim act

S1(2)
trim zero/gain

D/A trim, Scaled
D/A trim

Distributor
Model
Dev.id
Final asmbly num
Universal rev
Fld dev rev
Software rev
PV LSL
PV USL
Snsr1(2) LSL
Snsr1(2) USL
Snsr1(2) snsr
Term LSL
Term USL

Output

Display
(Note 3)

Monitoring

Maintenance

Adjustment

Referential
Information

Sensor burn-out
Output

Output in CPU
failure

Burst mode

Multi-drop mode

Display select

Display update
period

Process Variable

Output in %

Output in mA

Terminal
Temperature

Test Output

Self-diagnostics

Write Protect

Sensor trim
(Sensor1/2)

Analog output
trim

T0301_2.EPS

3-11

3-11

3-16

3-16

3-17

3-12

3-13

3-11

3-19

3-19

3-22

3-22

3-22

3-22

3-22

3-13

3-14

3-14

3-14

3-14

3-14

3-14

3-15

Note 3:

Appears only when Integral Indicator is specified.

Advertising
This manual is related to the following products: