6 tag no, 5 detailed setup, 1 device information – Yokogawa Wireless Temperature Transmitter YTA510 User Manual

Page 24: 2 test output, 3 burnout function, Tag no -11, Detailed setup -11, Device information -11, Test output -11, Burnout function -11

Advertising
background image

3. OPERATION

IM 01C50T01-01E

3-11

3.4.6 Tag No.

To change the Tag, see section 2.2.4 “Entering, setting
and Sending Data”.

Up to 8 characters can be set with “Tag”.

Example: To change from Tag “

YOKOGAWA

” to “

FIC-1A

.”

YTA :YOKOGAWA
Tag No.
YOKOGAWA
YOKOGAWA

HELP

DEL

ESC

ENTER

1. Device setup

3. Basic Setup

1. Tag

Enter new setting and press

ENTER[F4]

.

Call up the “

Tag

” setting display.

See Section 2.2.4.

3.5 Detailed Setup

3.5.1 Device Information

Following Device information can be entered.

Tag :

Up to 8 characters.

Tag Extension:

Extension of Tag description. Up
to 8 characters.

Descriptor:

Up to 16 characters

Message:

Up to 32 characters

Date:

mm/dd/yy mm:month, dd:day,
yy:year

Call up the “Device Information” display. [1.Device
setup

4.Detailed setup

4.Device information]

3.5.2 Test Output

This feature can be used to output a fixed current from
3.6 mA (-2.5%) to 21.6 mA (110%) for loop checks.

Call up the “Loop test ” display. [1.Device setup.

2.Diag/Service

2.Loop test]

Put the control loop in manual mode by pressing

OK[F4]

. The following output selections are offered.

4mA:

Outputs 4 mA current

20 mA:

Outputs 20 mA current

Other:

Sets a desired output in mA using

alphanumeric keys.

End:

Exits

CAUTION

1. Test output is held for approximately 10

minutes, and then released automatically
after the time has elapsed. If the HART
communicator power supply is turned off or
communication connector is disconnected
during the test output operation, it is held for
approximately 10 minutes.

2. Press the

key to release the test output

immediately.

NOTE

If the transmitter is equipped with the integral
indicator, the LCD displays F.O.

F0304.EPS

"F.O." lit on.

3.5.3 Burnout Function

a) Sensor burnout

Configure the burnout mode in the case of sensor
failure or disconnection.
Call up the “Snsr burnout type” display. [1.Device
setup

4.Detailed setup

3.Output condition

1.Analog output

5.Snsr burnout ]

When the sensor failure is detected, the transmitter
outputs one of the following values.
Low:

Outputs 3.6mA

High:

Outputs 21.6mA

User(mA):

Outputs user set value in mA. Settable
within 3.6 to 21.6 mA.

User(%):

Outputs user set value in %. Settable
within -2.5 to 110 %.

Off:

The burn out output is NOT defined

NOTE

When sensor burnout is set to “off”, the
transmitter's output will go undefined at sensor
failure. It is necessary to understand this point
well when setting “off”.

Advertising