3. OPERATION

IM 01C50T01-01E

3-12

During sensor burn out time, the Sensor1 failure or the
Sensor2 failure error message is generated. (See
Section 3.6.1 for details.)

If the transmitter is equipped with the integral indica-
tor, the LCD displays “Abn.” as shown in Figure 3.2.

F0306.EPS

Displays "Abn."

Figure 3.2 Integral Indicator Display in Sensor

Burn Out

b) Burnout in hardware failure

The output status of the transmitter in hardware
failure is set using a jumper on the CPU assembly.
The current setting can be checked in parameter
“AO Alrm type”.
Call up the “AO Alrm type parameter” display.
[1.Device setup

4.Detailed setup

3.Output

condition

1.Analog output

2.AO Alrm typ].

3.5.4 Integral Indicator Display Mode

When an integral indicator is specified with the
transmitter, the display items and update speed can be
configured.

Call up the”Meter output” display [1.Device setup

4.Detailed setup

3.Output condition

3.Meter

output ]

1) Display selection

(a) Process Disp

To Specify process variables to
be shown on the digital display.
Process variables and output
value(see (b)) are shown on the
digital display in turn. Select-
able from followings;

PV
SV
TV
4V
PV, SV
PV, SV, TV
PV, SV, TV, 4V
Inhibit

If two or more variables are
specified, each value appears
one after another in order. The
unit corresponding to each

process variable is lit. If
“Inhibit” is selected, no process
variable is displayed.

(b) % /mA Disp

To Specify output value to be
shown on the digital display.
Select from following;

mA
%
mA, %
Inhibit

If two values are selected, they
will appear on the display one
after another. Process variables
and output value(see (a)) is
shown on the digital display in
turn. The unit corresponding to
each output value is lit. If
“Inhibit” is selected, no ourput
value will be displayed.

(c) Err-No Disp

To Select whether error codes
are shown on the digital display
or not when an error occurs.
When “Show” is selected, error
code will be shown on the
digital display. When “Inhibit”
is selected, no error code will
be shown on the digital display.

(d) Bar graph

To select whether the output
bar-graph display is lit or not.
When “Show” is selected, The
output value is shown in a 32-
segment bar-graph. If “Inhibit”
is selected, the bar-graph is off.

(e) Matrix Disp

To select information to be
shown on the dot matrix
display.

Process:Display name of the

process variable shown on
the digital display. (Ex.
PV)

Type:

Display type of the input
sensor

Wire:

Display the number of
wires of the input sensor.

F03541

Displays PV value.

F03542

Displays output value.

F03543

Displays PV value and

sensor type

F03542

Displays bar graph.

Advertising
This manual is related to the following products: