3 review, 4 basic setup, 1 sensor configuration – Yokogawa Wireless Temperature Transmitter YTA510 User Manual

Page 18: Review -5, Basic setup -5, Sensor configuration -5

Advertising
background image

3. OPERATION

IM 01C50T01-01E

3-5

3.3 Review

Before starting operation, review all the configuration
of the transmitter to confirm that it meets the current
application.

YTA :
Review
1. Sensor 1 Review
2. Term temp Review
3. Out & Meter Review
4. Device Review

HELP

SAVE

HOME

ENTER

Call up “

Review

” display.

Parameters are grouped by type
and listed in review display of each
group.

Call up each review display, and
scroll through the list to check each
variable.
If a change is necessary, refer to the
“3.4 Basic Setup” in this manual.

1. Device setup

5. Review

3.4 Basic Setup

3.4.1 Sensor Configuration

When changing sensor type, it is necessary to change
the parameters related to the sensor type.

Figure 3.1 diagram shows the wire connections to the
input terminals of the transmitter and sensor type
selections for the parameters in each connection case.
Note that TCs and mV are categorized as Group A and
RTDs and ohm as Group B.

Check the connections between the input terminals and
temperature sensors and set the correct sensor type and
the number of wire connections for the parameters.

Thermocouple and DC voltage (TC & mV)

Resistance thermometer(RTD) and resistance (2-wire type)

1-input model YTA110, YTA310, and YTA320

2-input model YTA320

1
2
3
4
5

1
2
3
4
5

1
2
3
4
5

1
2
3
4
5

(–)

(+)

(B)

(A)

(B)

(B)

(A)

(B)

(A)

(A)

(B)

Thermocouple(TC) &
Resistance thermometer(RTD) and resistance (3-wire type)

1
2
3
4
5

(–)

(+)

(+)

F0303.EPS

Sensor1

Sensor1

Sensor1

Sensor1

1
2
3
4
5

(B1)

(A1)

(B2)

(A2)

Sensor2

Sensor1

1
2
3
4
5

(B1)

(B1)

(A1)

(B2)

(B2)

(A2)

Sensor2

Sensor1

Sensor2

Sensor1

1
2
3
4
5

(–)

(+)

(A)

(B)

(B)

Sensor2

Sensor1

Resistance thermometer(RTD) (4-wire type)

Resistance thermometer(RTD) and resistance (3-wire type)

Thermocouple and DC voltage (TC & mV)

Resistance thermometer(RTD) and resistance (2-wire type)

Resistance thermometer(RTD) and resistance (3-wire type)

Group A

Group B

Group B*

Group B

Group A

Group A

Group A

Group B

Group B

Group B

Group B

Group B

TYPE B (IEC584)
TYPE W3 (ASTM988)
TYPE W5 (ASTM988)
TYPE E (IEC584)
TYPE J (IEC584
TYPE K (IEC584)
TYPE L (DIN43710)
TYPE N (IEC584)
TYPE R (IEC584)
TYPE S (IEC584)
TYPE T (IEC584)
TYPE U (DIN43710)

Pt100 (IEC751)
Pt200 (IEC751)
Pt500 (IEC751)
JPt100 (JIS)
Ni120 (STI INC)
Cu (SAMA RC21-4)
ohm

mV

TYPE B (IEC584)
TYPE W3 (ASTM988)
TYPE W5 (ASTM988)
TYPE E (IEC584)
TYPE J (IEC584
TYPE K (IEC584)
TYPE L (DIN43710)
TYPE N (IEC584)
TYPE R (IEC584)
TYPE S (IEC584)
TYPE T (IEC584)
TYPE U (DIN43710)

Pt100 (IEC751)
Pt200 (IEC751)
Pt500 (IEC751)
JPt100 (JIS)
Ni120 (STI INC)
Cu (SAMA RC21-4)
ohm [

* Only for 2 or 3-wire type]

mV

Group A

Group B

* : Without ohm

Advertising