7 analog output trim, Analog output trim -15 – Yokogawa Wireless Temperature Transmitter YTA510 User Manual

Page 28

Advertising
background image

3. OPERATION

IM 01C50T01-01E

3-15

2. Call up the “Snsr1 inp trim” display. [1.Device

setup

2. Diag/Service

3.Calibration

3.Sensor Trim

1.Snsr 1 inp trim]

3. Select “2.Input Trimming Mode”. The following

selections are offered.

V.R. / ZERO&GAIN
V.R. / ZERO
Temp / ZERO&GAIN
Temp / ZERO

Select “V.R. / ZERO&GAIN” or “V.R. / ZERO”
when the calibration device is DC voltage generator
or Variable resistor or select “TEMP /
ZERO&GAIN” or “TEMP / ZERO” when the
device is Temperature sensor.

4. Enable the user trim. Select “3. Snsr1 Trim Act”.

The following selections are offered.

On:

Use User trim value

Off:

Ignore User trim value, and return to the
factory setting.

Clear: Clear User trim value and return to the

factory setting.

Select “On” to enable trim function and Press

ENTER[F4]

.

5. Perform zero-adjustment. Call up the “4. S1 Trim

Zero/Gain”.

(a) When the input trimming mode is “V.R. /

ZERO&GAIN” or “V.R. / ZERO”, apply the value
for the zero-point shown in the table 3.2 depending
on the specified sensor type. Wait until the input
from the calibration device becomes stable.

(b) When the input trimming mode is “TEMP /

ZERO&GAIN” or “TEMP / ZERO”, expose the
temperature sensor to calibration temperature for
the zero-point. Wait until the input from the
temperature sensor becomes stable.

Table 3.2

Sensor type

Zero-point value

TC, mV

RTD, ohm

B,R,S,T

E,J,K,N,W3,W5,L,U,mV

Pt100, JPt100, Ni120, Cu

Pt200, Pt500, ohm

0 mV

0 mV

40

40

Sensor type

Gain-point value

TC, mV

RTD, ohm

B,R,S,T

E,J,K,N,W3,W5,L,U,mV

Pt100, JPt100, Ni120, Cu

Pt200, Pt500, ohm

25 mV

75 mV

330

1600

T0302.EPS

6. Press

OK[F4]

.

7. Enter Zero trim reference value from key pad, and

press

ENTER[F4]

.

Perform the following steps when the input trim-
ming mode is “V.R. / ZERO&GAIN” or “Temp /
ZERO&GAIN”.

8. Perform gain-point adjustment.
(a) When the input trimming mode is “V.R. /

ZERO&GAIN”, apply the value for the gain-point
shown in the table 3.2 depending on the specified
sensor type. Wait until the input from the calibra-
tion device becomes stable.

(b) When the input trimming mode is “Temp /

ZERO&GAIN”, expose the temperature sensor to
calibration temperature for the gain-point. Wait
until the input from the temperature sensor becomes
stable.

9. Press

OK[F4]

.

10. Enter Gain trim reference value from key pad, and

press

ENTER[F4]

.

Follow outlined procedures to perform Sensor2 trim.

3.5.7 Analog Output Trim

Fine output adjustment is performed with “D/A trim”
or “Scaled D/A trim”.

D/A Trim

D/A trim is to be performed if the digital ammeter used
for calibration does not read 4.000mA and 20.000mA
exactly when the output signal is 0% and 100%.

Procedures

1. Call up the “D/A trim” display. [1.Device setup

2. Diag/Service

3.Calibration

4.AO D/A Trim

1.D/A trim]

2. Press

OK[F4]

to set the transmitter to manual

mode.

3. Connect the ammeter(

±

1

µ

A is measurable) and

press

OK[F4]

.

4. Press

OK[F4]

to make the transmitter output 4

mA(0%) . Check the reading of the ammeter, and
enter this value using the key pad and Press

ENTER[F4]

.

5. Next, press

OK[F4]

to make the transmitter output

20 mA (100%) . Check the reading on the ammeter,
and enter this value using the key pad.

6. After completing the above steps, return the

transmitter to automatic control.

Scaled D/A trim

Scaled D/A trim can be performed to adjust the output
when using a voltmeter or the other meters scaled to 0
to 100 %.

Advertising
This manual is related to the following products: