Yokogawa Wireless Temperature Transmitter YTA510 User Manual

Page 32

Advertising
background image

3. OPERATION

IM 01C50T01-01E

3-19

iimmediately. Warnings are the light errors and
indication of the status which is important for opera-
tion.

When an error message appears, see table 3.4 “List of
Errors” for details. If warning messages are set be
shown, the warning status will also be shown on the
HART communicator. See table 3.5 “List of Warnings”
for details.

YTA :
Test device
1 Status
2 Self test
3 Master test
4 Set warning enbl

HELP

SAVE

HOME

ENTER

1

YTA :
Self test OK

HELP

SAVE

ABORT

OK

2

1. Device setup

2. Diag/Service

1. Test device

YTA :
Test device
1 Status
2 Self test
3 Master test
4 Set warning enbl

HELP

SAVE

HOME

ENTER

3

YTA :
Status
1 Status group 1
2 Status group 2
3 Status group 3
4 Status group 4
5 Status group 5

HELP

SAVE

HOME

ENTER

4

OK

Call up the “

Test device

” display,

and select “

Self test

”.

If there is no error detected, “

Self

test OK

” will be displayed. When an

error occurs, an error message
appears, and the results of self-
diagnosis appears in the Status item.

Call up the “

1. Status

”.

The Error status is in groups 1 and
2.
The Warning status is in groups 3
through 6. Highlight the group to
show the status Errors. If Error is
not detected, the result of
diagnostics is indicated as “

OFF

”.

If “

ON

” is indicated, a

countermeasure for that error is
necessary.

Diagnostic by “

Self test

NOTE

Note that if the Set warning enbl is set to “OFF
(not display)”, the status in this display remains
“ OFF ” even when the warning status occurs.
(See 3.6.2)

Advertising
This manual is related to the following products: