<Appendix B Messages Summary>

B-1

IM 11B03G03-03E

Appendix B Messages Summary

This appendix lists the error messages that may appear while you are working with the Mainte-

nance Terminal, along with the corrective measures.

n

Analyzer Operation Window

Message

Description

Corrective Measures

Common

Stop command has been

received.

The communication link has

been dropped because the

analyzer panel is being operated

upon in the field.

After completing the analyzer

operation in the field, execute the

Connection command.

Exit

Close Maintenance

Terminal after Analysis

Results Display is closed.

Invalid attempt to exit the

Maintenance Terminal without

closing the analysis results

window.

Close the analysis results

window before exiting the

Maintenance Terminal.

Analyzer

operation

Analyzer has not

accepted operation

command.

Invalid attempt to change the

operation mode or operate

valves while at a wrong user

level.

Change the mode or operate

valves at user level C only.

Valve operation is only

available in Manual

mode.

Invalid attempt to operate valves

in a mode other than the Manual

mode.

Select the Manual mode when

operating valves. However, note

that valves are operational at

user level C only.

Detector operation is

only available in Manual

mode.

Invalid attempt to operate the

detector in a mode other than the

Manual mode.

Select the Manual mode when

operating the detector. However,

note that the detector is

operational at user level C only.

A higher user level is

needed for pattern/ valve

/detector operation.

Invalid attempt to change the

operation pattern or operate

valves at a wrong user level.

The operation pattern can be

changed at user levels B and C

only; valves can be operated at

user level C only.

File

operation

File access is only

available at user level C.

Invalid attempt to upload or

download parameter settings at

the wrong user level.

Parameter settings can be

uploaded and downloaded at

user level C only.

File operation is only

available in Process/Lab

- Stop.

Invalid attempt to upload or

download parameter settings in

a mode other than Process/Lab

- Stop.

Make Process/Lab - Stop active

before uploading or downloading

parameter settings.

n

Chromatogram Window

Message

Description

Corrective Measures

Continuous

Saving

There is less than 1MB

of free space on the hard

disk.

The free space on the hard

disk has fallen below 1 MB

during the continuous saving of

chromatograms.

Delete unnecessary files to make

an free space of 1 MB or mode.

Pen

selection

This file is not a

Chromatogram file.

Invalid attempt to open a file

other than a chromatogram file.

Select the correct file type.

This function is already in

use by another pen.

The selected differential

chromatogram is already active

for another pen.

Select a different chromatogram

or delete the currently active

chromatogram.

4th Edition : Mar. 21, 2012-00