Ndetailed display of temperature controller, Ldescriptions, Lprocedure – Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual

Page 39

Advertising
background image

<3. Analyzer Operation Window>

3-10

IM 11B03G03-03E

l

Descriptions

The following table summarizes the descriptions of the display.

Item

Available Information

Type

Indicates the type of detector. The window shows TCD, FID, FI-CONV (FID with

methane converter), FPD or None.

Input signal

Indicates the value obtained by averaging the A-D converted analog signal of the

detector sampled every 20 msec by the value set in the sample rate.

Filtered signal

Indicates the value obtained by filtering the input signal by the value set in the filter

constant in the detector signal setting.

Standard deviation

Indicates the standard deviation of the output signals of the last 20 analysis.

Voltage

Indicates the bridge current of the TCD (TCD only).

Current

Indicates the bridge current of the TCD (TCD only).

n

Detailed Display of Temperature Controller

l

Procedure

(1) Place the cursor on the temperature controller symbol in the Valve Display/Control area and

click the left mouse button. The Temperature Control Unit Operation window opens.

F0314.ai

Temperature

status detailed

display window

Click left

mouse button

Valve display and operation area

[Close]

command

(2) To close the Temperature Control Unit Operation window, click on the [Close] command in

the control menu.

l

Window Layout (Temperature Control Unit Operation Window)

F0315.ai

4th Edition : Mar. 21, 2012-00