Nuploading, Lprocedure, Help – Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual

Page 50

Advertising
background image

<3. Analyzer Operation Window>

3-21

IM 11B03G03-03E

n

Uploading

To upload and save parameter settings from the analyzer to a disk of the Maintenance Terminal,

take the following procedure.

l

Procedure

(1) Click on the [Upload to PC] command in the [File] menu. The Save As dialog box pops up

where you can specify the file name for the parameter setting information to be uploaded.

The default directory is “Data” right under the installation directory.

Input

F0336.ai

(2) Type the name of the file where you want to save the parameter settings, with an extension

of “.pr6” in the “File name” field.

Example: When the file name is “system01”, type

system01.pr6

(3) Click on the [OK] button. A dialog box pops up to show the status of the transmission of the

parameter settings from the analyzer.

F0337.ai

When all parameter settings have been uploaded and saved in the file, the following dialog

box pop up.

F0338.ai

• It takes about 40 seconds for data transmission from the analyzer (if the communication

speed is 38400 bps).

(4) Click on the [OK] button. The uploading has been completed.

HELP !

To interrupt the uploading of parameter settings;

• Click on the [Interrupt] button in the dialog box showing the transmission status. A confirmation dialog box

pops up.

• Click on the [Yes] button.

4th Edition : Mar. 21, 2012-00

Advertising