Ndeleting a user alarm description – Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual

Page 75

Advertising
background image

<5. Alarm Windows>

5-8

IM 11B03G03-03E

(5) Repeat steps 2 through 4 for defining descriptions for another user-defined alarms, if any.
(6) Click on the [Close] button to finish.

n

Deleting a User Alarm Description

l

Procedure

(1) In the Alarm Status or Alarm History window, click on the [Making User Alarm Description]

command in the [Detail] menu.

(2) Type in the alarm number and click on the [Display] button. The defined description for the

alarm number appears.

(3) Click on the [Delete] button. A confirmation dialog box pops up. If you want delete the on-

screen user alarm description, click on the [OK] button.

(4) The user alarm description of the specified alarm number is erased and deleted from

initialization file “GIMT.INI”.

(5) Click on the [Close] button to finish.

4th Edition : Mar. 21, 2012-00