Ntiming display area, Nvalve display/control area, Ldescriptions of the area – Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual

Page 36: Larea layout

Advertising
background image

<3. Analyzer Operation Window>

3-7

IM 11B03G03-03E

l

Descriptions of the Area

The scales are displayed in the following format:

Item

Format

Horizontal scale

Shows a chromatogram with the analysis period set on a task-by-task basis.

Vertical scale

An automatic scale, which adjusts automatically so that the approximate data

deflection up to that moment is 80 percent of full scale. This automatic scaling is set

separately for the chromatogram and temperature.

Active Stream

Indicates the number of the stream being currently measured. Indicates "Disp.OK"

when sample displacement is ready (available only for manual calibration/validation

method).

Next Stream

Indicates the number of stream scheduled next. For preparation status, (P) is added as

a suffix.

n

Timing Display Area

This area displays a chart of the on/off times for each analyzer valve and gate according to the

analysis cycle. These times are preset. As for the on/off times for the gate, if two detectors are

used, this area shows the times separately for each detector.

l

Area Layout

F0309.ai

Indicates time elapsed.

Indicates ON status.

Gate (one detector)

or

Gat1/Gat2 (two detectors)

Valve 1 to 8

l

Descriptions of the Area

The following table summarizes the descriptions of the display.

Item

Available Information

Horizontal scale

Indicates the times of an analysis cycle set on a stream-by-stream basis.

Valve 1 to 8

Indicates on/off states of valves 1 to 8 in bar graph.

Gate or Gat1/Gat2

Indicates on/off state(s) of gate(s) in bar graph.

n

Valve Display/Control Area

The Valve Display/Control area displays the state of the stream switching valve of the analyzer

(the currently active valve numbers or OFF), the on/off state of the analyzer valves and detectors,

and the state of the temperature controller. In this area, these states can be changed.

TIP

In order to actually operate valves, the system must be at user level C as well as in the Manual mode.

4th Edition : Mar. 21, 2012-00