Lsignal terminal, to the gc1000 mark ii, Lsignal terminals on the communication equipment, Lpower supply – Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual

Page 16: Lpc communication cable (straight type)

Advertising
background image

<Installing the Maintenance Terminal>

9

IM 11B03G03-03E

n

Standard Specifications (Communication Converter K9404LA,

K9490LD, RS-232C, Communication cable, Figure, Wiring)

This communication converter converts communications from RS-422 to RS-232C for PC

or DCS communication and shuts off the communication signals when a power-off signal is

received from the analyzer. This maintains explosion-protection on the analyzer side.

l

Signal terminal, to the GC1000 Mark II

M4 terminals
1

RS-422 SDA, SDB, RDA, RDB

2

D/I 12 V +, -

l

Signal Terminals on the Communication Equipment

M4 terminals
RS-232C D-Sub25 female terminals

l

Power supply

100 to 240 V AC ±10%, 50/60 Hz ±5%, 15 W Max., M4 terminals

l

Terminal Arrangement (common for K9404LA, K9404LD)

RS-232C

SDA SDB RDA RDB

RS-422

D/I 12V

+

N

L

CN1

CN2

CN3

D-SUB female 25 pins

Terminal: M4

from the GC1000 analyzer

AC power supply

F10.EPS

l

PC Communication Cable (Straight type)

GND

1
8
3
2

20

7
6
4
5

22

FG

CD
RD

SD
ER

SG
DR

RS
CS

CI

GND

1
2
3
4
5
6
7
8
9

CD
RD
SD
ER
SG
DR
RS
CS
CI

Internal wiring

K9404LA

Personal computer

F12.ai

4th Edition : Mar. 21, 2012-00

Advertising