<6. Analysis Results Windows>

6-18

IM 11B03G03-03E

6.7

Resetting Historical Data

If you want to record data from scratch in the Concentration Analysis History and Retention Time

History windows, execute the [Reset History] command. Executing the command simultaneously

erases all data in both windows, regardless of the state of the analysis results windows.

n

Resetting Historical Data in the Concentration Analysis History and

Retention Time History Windows

l

Procedure

(1) Open the Latest Analysis Results window and click on the [Reset History] command in the

[File] menu. A dialog box pops up to ask you if you want to reset the historical data.

F0618.ai

F0619.ai

(2) Click on the [OK] button. This erases all data in the Concentration Analysis History and

Retention Time History windows.

The heading “Strm#-Peak#-Peak Name” remains in the windows.

NOTE

If you have made any change to the operation pattern or peak settings while the Analysis Results

window is active:

• Execute the [Exit] command in the [File] menu to close the Analysis Results window once.

Next, exit the Maintenance Terminal. And then restart the Maintenance Terminal, and

reopen the Analysis Results window. Executing the [Reset History] command without exiting

the Analysis Results window results in incorrect data display.

4th Edition : Mar. 21, 2012-00