<3. Analyzer Operation Window>

3-2

IM 11B03G03-03E

3.1

Starting and Exiting the Maintenance

Terminal

This section describes how to start and exit the Maintenance Terminal.

n

Starting

While the LCD emulator is running, you cannot start the Maintenance Terminal. Exit the LCD

emulator before you start the Maintenance Terminal. You also cannot start the Maintenance

Terminal unless the analyzer is in the Remote mode. Note that there is only one window from

which the Maintenance Terminal can be started.

l

Procedure

(1) Turn on the power to the personal computer, start up Windows™, and open the

Maintenance Terminal group window of Program Manager.

F0302.ai

Maintenance

Terminal icon

(2) Double-click on the Maintenance Terminal icon.

This establishes a communication link with the analyzer and the Initial Database Loading

window opens.

F0303.ai

Initial database

loading window

Initial database

loading dialog

Analyzer communication task icon