App index digital display – Yokogawa Button Operated DX2000 User Manual

Page 26

Advertising
background image

1-15

IM 04L42B01-01E

Overview of Functions

1

2

3

4

5

6

7

8

9

10

11

12

13

App

Index

Digital Display

Displays the measured data numerically using large numbers. For the operating
procedure, see section 4.2.

Unit

Tag or channel numbers

Alarm mark

Measured value

Note

• Numeric display of measurement channels

If a measured value of a measurement channel is over range (see below), the measured

value is indicated as “+Over” or “–Over.” If a burnout is detected on a channel whose

burnout detection function is enabled, the word “Burnout” is indicated. Otherwise, a numeric

value is displayed.

Over range of measurement channels

• For DC voltage input, over range occurs when the measured value of the measurement

channel exceeds ±5% of the measurable range. For example, the measurable range when

the measurement range is 2 V is –2.000 to 2.000 V. If the measured value exceeds 2.200 V,

+ over range occurs; if the measured value falls below –2.200 V, – over range occurs.

• For thermocouple or RTD input, over range occurs when the measured value exceeds

approximately ±10°C of the measurable range. For example, the measurable range

when the measurement range is R is 0.0 to 1760.0°C. If the measured value exceeds

approximately 1770.0°C, + over range occurs; if the measured value falls below

approximately –10.0°C, – over range occurs.

• For channels that are linearly scaled, + over range occurs when the value exceeds 30000

excluding the decimal point; – over range occurs when the value falls below –30000.

However, + over range can be changed to greater than or equal to 105% of the scale width
and – over range to less than or equal to –5% of the scale width within ±30000.

For the

setting procedure, see section 3.11.

• Numeric display of computation channels

See section 1.8, “Computation and Report Function (/M1 and /PM1 Options)

• Numeric display of external input channels (/MC1 option)

The numeric range that can be displayed is –30000 to 30000 excluding the decimal point.

The decimal place corresponds to the decimal place of the lower limit of span of the external

input channel. On the numeric display, values are displayed if the value is within the –30000

to 30000 range regardless of the upper and lower limits of span.

If the value exceeds 30000, + over range occurs; if the value falls below –30000, – over

range occurs.

• Alarm Indication

Alarm mark and measured value are displayed as follows according to the alarm

status. When you use the alarm annunciator function (release number 3 or later), the
alarm mark follows the annunciator sequence.

Alarm

Release

Alarm ACK

Alarm ACK

Occurrence

When indication is

set to non-hold

When indication

is set to hold

Red

Blue

Red

Blue

Red

Red

Blue

Blue

Blue

Blue

Blue

Measured

value

Blinking

red

Blinking

green

Blinking

red

Green

Red

Red

Green

Green

Green

Green

Green

Alarm mark

1.3 Display

Advertising
This manual is related to the following products: