DC/AC 24 V Power Supply (/P1)

Item

Specifications

Rated supply voltage

24 VDC and 24 VAC (50/60Hz)

Allowable power supply voltage range

21.6V to 26.4 VDC/AC

Insulation resistance

Between power terminal and earth: 20 MΩor greater at 500 VDC.

Withstand voltage

Between power terminal and earth: 500 VAC at 50/60 Hz for one minute

Rated power supply frequency (for AC)

50/60 Hz

Allowable power supply frequency range (for AC)

50 Hz±2%, 60 Hz±2%

Power supply fluctuation (for AC)

With variation within 21.6 to 26.4 VDC/AC: ±1digit or less
With variation of ±2 Hz from rated power supply frequency (at 24 VAC): ±(0.1% of rdg+1digit) or

less

Rated power consumption

45 VA (for DC), 70 VA (for AC)

Power consumption

Supply voltage

LCD backlight off

Normal

Maximum

24 VDC

12 VA

20 VA

45 VA

24 VAC (50/60Hz)

20 VA

34 VA

70 VA

Multi Batch Function (/BT2; release number 3 or later)

Item

Specifications

Function

Individual recording start/stop and file generation operations can be performed for each batch.
Equippable models: DX2010, DX2020, DX2030, DX2040, and DX2048

Number of batches

2 to 12

Batch single operations

Recording start/stop, computation reset, message write

Batch overview operations

Computation start/stop, report start/stop, manual sampling, setup data save/load

Scan interval

Maximum speed of 1 s in normal mode only (same for all batches)

Data type

Display or event data only. Trigger mode cannot be used for event data.

Data recording interval

Same for all batches

Data files

Can be displayed or made into an event data file separately for each batch

Number of display groups

Up to twelve per batch

The maximum number of channels per group is ten.

Channels in a display group are sampled for display or event data.

Timers and match time timers 12 max.
Batch single settings

Group, trip line, file header, data file name, text field, batch number, and lot number

PROFIBUS-DP (/CP1 option; release number 3 or later)

Item

Specifications

Function

A PROFIBUS-DP master device can access internal data in the following ways:
Load measurement channel data
Load computation channel data
Write to communication input data (for up to 32 channels)

Data mapping

Buffer

Description

Maximum Size

Input

Measurement channels are arranged at the start of the buffer. As

many computation channels as can be placed into the buffer are

inserted after the measurement channels.

128 bytes

Output

Up to 32 communication input data values is arranged.

128 bytes

Node address

0 to 125

Interface

PROFIBUS-DP-V0 Slave

Transmission medium

Two dedicated cables (one for each signal line)

Transmission speed/distance 9.6 kbps/1200 m to 12 Mbps/100 m
Terminator

No internal terminator (must be terminated externally)