App index standard performance – Yokogawa Button Operated DX2000 User Manual

Page 339

Advertising
background image

13-21

IM 04L42B01-01E

Specifications

1

2

3

4

5

6

7

8

9

10

11

12

13

App

Index

Standard Performance

Item

Specifications

Measurement/display accuracy

Standard operating conditions:

Temperature:

23 ± 2°C

Humidity:

55% ± 10%RH

Power supply voltage:

90 to 132 or 180 to 250 VAC

Power supply frequency:

50/60 Hz ± 1%

Warm-up time:

At least 30 minutes.

Other ambient conditions such as vibration should not adversely affect the operation.

Input Type

20 mV
60 mV

200 mV

2 V

1-5 V

6 V

20 V
50 V

R

S

B

K

E

J

T

N

W

L

U

WRe

Pt100

JPt100

Voltage

Contact

A/D integration time: 16.7 ms or more
±(0.05% of rdg + 12 digits)

±(0.05% of rdg + 3 digits)

±(0.05% of rdg + 12 digits)

±(0.05% of rdg + 3 digits)

±(0.15% of rdg + 1°C)
R, S

0 to 100°C: ±3.7°C,

100 to 300°C: ±1.5°C

B

400 to 600°C: ±2°C,

Accuracy not guaranteed for

values less than 400°C

±(0.15% of rdg + 0.7°C)
–200 to –100°C: ±(0.15% of rdg + 1°C)
±(0.15% of rdg + 0.5°C)
–200 to –100°C: ±(0.15% of rdg + 0.7°C)

±(0.15% of rdg + 0.7°C)
–200 to 0°C: ±(0.35% of rdg + 0.7°C)
Accuracy not guaranteed for values less
than –200°C.
±(0.15% of rdg + 1°C)
±(0.15% of rdg + 0.5°C)
–200 to –100°C: ±(0.15% of rdg + 0.7°C)
±(0.2% of rdg + 2.5°C)
0 to 200°C: ±4.0°C

±(0.15% of rdg + 0.3°C)

Threshold level (Vth=2.4 V) accuracy ± 0.1 V
1 kΩ or less: 1 (ON), 100 kΩ or more: 0 (OFF) (parallel capacitance of 0.01 μF or less)

A/D integration time: 1.67 ms
±(0.1% of rdg + 40 digits)

±(0.1% of rdg + 15 digits)

±(0.1% of rdg + 40 digits)

±(0.1% of rdg + 15 digits)

±(0.2% of rdg + 4°C)
R, S

0 to 100°C: ±10°C,

100 to 300°C: ±5°C

B

400 to 600°C: ±7°C,

Accuracy not guaranteed for

values less than 400°C

±(0.2% of rdg + 3.5°C)
–200 to –100°C: ±(0.15% of rdg + 6°C)
±(0.2% of rdg + 2.5°C)
–200 to –100°C: ±(0.2% of rdg + 5°C)

±(0.3% of rdg + 3.5°C)
–200 to 0°C: ±(0.35% of rdg + 0.7°C)
Accuracy not guaranteed for values less
than –200°C.
±(0.3% of rdg + 7°C)
±(0.2% of rdg + 2.5°C)
–200 to –100°C: ±(0.2% of rdg + 5°C)
±(0.3% of rdg + 10°C)
0 to 200°C: ±18.0°C

±(0.3% of rdg + 1.5°C)

DC voltage

Thermocouple

• Not including

the accuracy

of reference

junction

compensation

• With burnout

detection

function OFF

RTD

DI

Max. Resolution of

Digital Display

1 µV

10 µV
10 µV

100 µV

1 mV
1 mV
1 mV

10 mV

0.1°C

Measurement Accuracy (Digital Display)

Range

13.6 General Specifications

Advertising
This manual is related to the following products: