13-1

IM 04L42B01-01E

Specifications

1

2

3

4

5

6

7

8

9

10

11

12

13

App

Index

13.1 Signal Input and Alarm

Measurement Input

Item

Specifications

Number of inputs, scan interval, and A/D integration time

DX2004
DX2008
DX2010
DX2020
DX2030
DX2040
DX2048

25 ms

125 ms

4
8
10
20
30
40
48

Fast sampling mode*

Normal mode

Number of inputs

Model

Scan interval

Integration time of the A/D converter

60 Hz/50 Hz/100 ms

60 Hz/50 Hz

600 Hz (fixed)

125 ms, 250 ms

1 s, 2 s, 5 s

2 s, 5 s

* Not available on models equipped with external input channels (/MC1 option) or when the multi

batch function (/BT2 option) is being used.

Input Type

DC voltage, 1-5V, thermocouple (TC), resistance temperature detector (RTD), ON/OFF input (DI),

and DC current (by adding an external shut resistor)

Measurement range and measurable range

DC voltage

20 mV

–20.000 to 20.000 mV

60 mV

–60.00 to 60.00 mV

200 mV

–200.00 to 200.00 mV

2 V

–2.0000 to 2.0000 V

6 V

–6.000 to 6.000 V

20 V

–20.000 to 20.000 V

50 V

–50.00 to 50.00 V

1-5V

1 to 5 V*

6

0.800 to 5.200 V

Thermocouple R*

1

0.0 to 1760.0°C

S*

1

0.0 to 1760.0°C

B*

1

0.0 to 1820.0°C

K*

1

–200.0 to 1370.0°C

E*

1

–200.0 to 800.0°C

J*

1

–200.0 to 1100.0°C

T*

1

–200.0 to 400.0°C

N*

1

–270.0 to 1300.0°C, on models

with release numbers 3 and later.

0.0 to 1300.0°C, on models with

release numbers 2 and earlier.

W*

2

0.0 to 2315.0°C

L*

3

–200.0 to 900.0°C

U*

3

–200.0 to 400.0°C

WRe*

4

0.0 to 2400.0°C

RTD

Pt (Pt100)*

5

–200.0 to 600.0°C

JPt (JPt100)*

5

–200.0 to 550.0°C

DI

Level

0: Less than 2.4 V. 1: 2.4 V or higher (judged at the 6 V range)

Contact*

7

0: Open. 1: Closed (parallel capacitance of 0.01 µF or less)

*1: R, S, B, K, E, J, T, N: IEC584-1 (1995), DIN IEC584, JIS C1602-1995

*2: W: W-5%Re/W-26%Re (Hoskins Mfg. Co.), ASTM E988

*3: L: Fe-CuNi, DIN43710, U: Cu-CuNi, DIN43710

*4: WRe: W-3%Re/W-25%Re (Hoskins Mfg. Co.)

*5: Pt100: JIS C1604-1997, IEC751-1995, DIN IEC751-1996

JPt100: JIS C1604-1989, JIS C1606-1989

Measuring current: i = 1mA (Pt100, JPt100)

*6: The range for linear scaling of 1-5V inputs. Burnout detection and low-cut

functions are available.

*7: The detected current value is approx. 10 μA.

32 to 3200°F

32 to 3200°F

32 to 3308°F

–328 to 2498°F

–328.0 to 1472.0°F

–328.0 to 2012.0°F

–328.0 to 752.0°F

–454 to 2372°F

, on models with

release numbers 3 and later.

32 to 2372°F

, on models with

release numbers 2 and earlier.

32 to 4199°F

–328.0 to 1652.0°F

–328.0 to 752.0°F

32 to 4352°F

–328.0 to 1112.0°F

–328.0 to 1022.0°F

Measurable Range

Range

Input Type