6 control (zone, graph (, partial)), 7 math (color), 8 math (zone, graph (, partial)) – Yokogawa Data Acquisition with PID Control CX2000 User Manual

Page 58: 5 save/load, clear data

Advertising
background image

58

IM 04L31A01-02E

List of Parameters

<#6 Control (Zone, Graph (, Partial))>

Parameter

Selectable Range or Selections

Initial Value

First-CH

101 to 118 or 201 to 248

101

Last-CH

101 to 118 or 201 to 248

101

Zone Lower

0% to 95%

0%

Zone Upper

5% to 100%

100%

Division

4, 5, 6, 7, 8, 9, 10, 11, 12, or C10

10

Bar graph

Normal or Center

Normal

Scale position

Off or 1 to 10

1

Partial

On/Off

Off

Expand

1% to 99%

50%

Boundary

Minimum span value + 1 digit to maximum span value – 1 digit or minimum

0.1

scale value + 1 digit to maximum scale value – 1 digit (when the channel is set
to scaling or square root computation)

<#7 Math (Color)>

Parameter

Selectable Range or Selections

Initial Value

CH31, CH41, CH51

Red, green, blue, blue violet, brown, orange,

Red

yellow-green, light blue, violet, gray,
lime, cyan, dark blue, yellow, light gray, or purple

CH32, CH42, CH52

Same as above

Green

CH33, CH43, CH53

Same as above

Blue

CH34, CH44, CH54

Same as above

Blue-violet

CH35, CH45, CH55

Same as above

Brown

CH36, CH46, CH56

Same as above

Orange

CH37, CH47, CH57

Same as above

Yellow-green

CH38, CH48, CH58

Same as above

Light blue

CH39, CH49, CH59

Same as above

Violet

CH40, CH50, CH60

Same as above

Gray

<#8 Math (Zone, Graph (, Partial))>

Parameter

Selectable Range or Selections

Initial Value

First-CH

31 to 60

31

Last-CH

31 to 60

31

Zone > Lower

0% to 95%

0%

Zone > Upper

5% to 100%

100%

Graph > Division

4, 5, 6, 7, 8, 9, 10, 11, 12, or C10

10

Graph > Bar graph

Normal or Center

Normal

Graph > Scale position

Off or 1 to 10

1

Partial

On/Off

Off

Expand

1% to 99%

50%

Boundary

Minimum span value + 1 digit to maximum span value – 1 digit or minimum

0.00

scale value + 1 digit to maximum scale value – 1 digit (when the channel is set
to scaling or square root computation)

#5 Save/Load, Clear data

Parameter

Selectable Range or Selections

Initial Value

#1 Save settings

#2 Load settings

#3 Save data

#4 Load display data

#5 Load event data

#6 File list

#7 Delete

#8 Format

#9 Clear data

Advertising