After making the connections according to the explanation given on the calibration

execution screen 2, click the Calibrate button.

The figure below is an example of the 4-CH, High-Speed Universal Input Module.

Exp. Calbraton executon screen 2

Start calibration

The figure below is an example of the 8-CH, Medium-Speed Analog Output
Module.

Exp. Calbraton executon screen 2

Start calibration

Click the Calibrate button.

The message, “Calibrating. Please wait and do not perform any other operation.” appears.

5.

Repeat steps 1 to 4 until all calibrations are completed.

6.

From the File menu, choose Write.

The calibration values are written to the input module.

Note

During calibration, do not perform any other operations (especially turning the MX100 OFF or
dropping communications). Doing so may cause the MX100 main unit to stop working.