5.2 Error Messages and Ther Correctve Actons

Error Messages on the Integraton Montor

Error Code

Message

Correctve Acton/Descrpton

Reference

Page

IP address is invalid.

Enter the correct IP address.

2-7

Subnet mask is invalid.

Enter the correct subnet mask.

2-7

Default gateway is invalid.

Enter the correct gateway address.

2-7

Failed to update.

Attempted to change the IP address while the connection is

established, or the MX100 you wish to configure is connected

by another software program. Disconnect the MX100.

-

Search MX100s nearby again.

Click the Search button again on the Connection screen, and

change the settings.

2-1, 2-5

Broken modules exist.

Remove the input/output modules one by one and determine

the broken module, then replace it.

*

Some modules are not calibrated

correctly.

Recalibrate.

4-4

Undetermined directory.

The save destination of the record file cannot be found. Check

whether the save destination exists.

2-16

Write error

The record file cannot be created. Check whether free space

on the destination drive exists, whether write permissions have

been granted, whether the disk is damaged, and whetherthe

file system is corrupt.

-

E5102

Failed to format CompactFlash.

Check whether communication with the MX100 is being

performed.

Insert a CF card correctly.

2-7

*

E5103

Invalid serial number. Please

reinstall the software.

Enter the correct serial number.

1-5

E5104

Failed to create folder.

The directory of the folder to create is not correct. Check the

file save destination.

2-16

E5105

Invalid folder name. Please enter

a new name.

Entered folder name is incorrect. Enter a correct folder name.

-

E5106

Cannot contain any of the

following characters:

\\ / : , ; * ? " < > | CON PRN AUX

CLOCK$ NUL COM1-9 LPT1-9

Entered folder name includes illegal characters. Enter a correct

folder name.

-

E5107

Compact Flash slot is empty.

Insert a CF card correctly.

*

W5131

Update the display settings.

The display setup is initialized according to the measurement

and computation channel settings.

2-43

W5132

Writing data files Currently.

During finalization, do not close MX100 Standard.

2-51

W5133

Cancel the current settings and

create a new project?

Settings are discarded and the MX100 Standard is restored to

its default state.

2-6

W5134

The current channel setup

contains invalid use of reference

channels.

Click No and change the settings of the improper channels. To

disable the channels with improper settings, click Yes.

2-25 to 2-27

W5135

In order to balance the strain

channels, it's necessary for

channels to be set up correctly.

The current channel setup

contains invalid use of reference

channels.

Click No and change the settings of improper channels. To

disable the channels with improper settings, click Yes.

2-25 to 2-27

W5136

Format Compact Flash?

The CF card is formatted.

2-11

W5137

Stop Record?

Data saving stops.

2-51

W5138

Module settings which do not

match the current software

configuration will be initialized.

If you click Yes, settings of modules that differ between the

MX100 module configuration and the configuration recognized

by MX100 Standard will be initialized. To not reconfigure, click

No.

2-10

W5139

Some detected modules do

not match the current software

configuration. Rebuild based on

the current MX hardware setup?

Click Yes reconfigure. To not reconfigure, click No.

2-9

* See the MX100 Data Acquisition Unit User’s Manual (IM MX100-01E).