2.4 Settng the Measurement Condtons (Settng

the Measurement Channels)

Procedure

Turn ON/OFF collectively

*3

When making the settings

the same as the first item

*3

Reset to default

*3

To copy

*3

*3 Applied to the selected range by dragging the cursor.

Applied to all lines if a range is not selected.

*1 Excluding Tag No. and Tag comment

*2 Cannot be used when Mode is set to RRJC

Record check box

Icon of the attached modules

Point to the icon to display information

about the module

Click the icon to select the channels of

the module

(gray background and red characters)

Monitor check box
Select the input mode

If changed, the Range setting and

others to its right are initialized

*1

Select the measurement range

If changed, the Span setting and

others to its right are initialized

*1

Minimum/Maximum value of span
The Use check box for Scale

*2

Decimal point position of the scale

*2

Maximum/Minimum values of the scale

*2

Click here

Measurement channel tab

1.

Click Channel to show the Channel setup screen.

2.

Click Meas. Channel tab.

Settng the Measurement Channels for Recordng and Montorng

3.

Select Record check boxes of the channels to record.

4.

Select the Monitor check boxes of the channels to monitor.

Settng the Input Mode

5.

Select the input mode from the Mode list box.

Settng the Measurement Range

6.

Select the measurement range from the Range list box.

Settng the Measurement Span

7.

Click the Min box or Max box under Span and enter the minimum or the maximum

value of the span.

Advertising