App index, Caution – Yokogawa µR20000 User Manual

Page 33

Advertising
background image

3-5

IM 04P01B01-17E

Using

the
RS-422A/485

Communication

Interface

(/C3
Option)

1

2

3

4

5

6

App

Index

(The following figure illustrates the case when the host computer’s interface is RS-232.)

Terminator (externally attached) 120 Ω, 1/2 W or more

#1

Do not connect terminator to #1 to #n–1

#2

#n

(#n ≤ 31)

Terminator (externally attached)

FG

SG

RD B

RD A

SD B

SD A

(SG)

(B)

(A)

(B)

(A)

FG

SG

RD B

RD A

SD B

SD A

(SG)

FG

SG

RD B

RD A

SD B

SD A

(SG)

(B)

(A)

SHIELD

TD( + )

TD( – )

RD( + )

RD( – )

Host

computer

Converter

RS-232

RS-422A/485

terminal on the

Recorder

Note

• The method used to eliminate noise varies depending on the situation. In the connection

example, the shield of the cable is connected only to the recorder’s ground (one-sided
grounding). This is effective when there is a difference in the electric potential between
the computer’s ground and the recorder’s ground. This may be the case for long distance
communications. If there is no difference in the electric potential between the computer’s
ground and the recorder’s ground, the method of connecting the shield also to the
computer’s ground may be effective (two-sided grounding). In addition, in some cases, using
two-sided grounding with a capacitor connected in series on one side is effective. Consider
these possibilities to eliminate noise.

• When using the two-wire type interface (Modbus protocol), the 485 driver must be set to

high impedance within 3.5 characters after the last data byte is sent by the host computer.

• Serial Interface Converter

Recommended converter
SYSMEX RA CO., LTD./RC-770X, LINEEYE CO., LTD./SI-30FA, YOKOGAWA/ML2

CAUTION

Some converters not recommended by Yokogawa have FG and SG pins that are
not isolated. In this case, do not connect anything to the converter’s FG and SG
pins (unlike the figure on the previous page). Especially in the case of long distance
communications, the potential difference that appears may damage the recorder
or cause communication errors. For converters that do not have the SG pin, they
can be used without using the signal ground. For details, see the manual that came
with the converter.

On some non-recommended converters, the signal polarity may be reversed (A/B or

+/– indication). In this case, reverse the connection.

3.2 Terminal Arrangement and Signal Names and the Connection Procedure of the RS-422A/485 Communication Interface