Chapter 4 commands, 1 command syntax, 1 command syntax -1 – Yokogawa µR20000 User Manual

Page 44: App index

Advertising
background image

4-1

IM 04P01B01-17E

Commands

1

2

3

4

5

6

App

Index

4.1 Command Syntax

The syntax of the setting/basic setting/output commands (see sections 4.4 to 4.7) of the
instrument is given below. ASCII codes (see appendix 1) are used for the character codes.
For the Maintenance/Test command syntax, see section 4.9.
For the Instrument Information server command syntax, see section 4.10.

Command name

Parameter

?

Terminator

Delimiter
Sub delimiter

SR02,SKIP;SR 03,VOLT,2V,-1500,1800

Parameter (

02 SKIP·····1800)

Delimiter (

,)

Command name (

SR)

Sub delimiter (

;)

Command example

• Command Name

Defined using two alphabet characters.

• Parameter

• Command parameters.
• Set using alphabet characters or numerical values.
• Parameters are separated by delimiters (commas).
• When the parameter is a numerical value, the valid range varies depending on the

command.

• Spaces before and after of the parameter are ignored (except for parameters that

are specified using an ASCII character string (unit, tag, and message string), when
spaces are valid.)

• You can omit the parameters that do not need to be changed from their current

settings. However, delimiters cannot be omitted.

Example SR01,,2V<terminator>

If multiple parameters are omitted and delimiters occur at the end of the command,

those delimiters can be omitted.

Example SR01,VOLT,,,<terminator> → SR 01,VOLT<terminator>

• The number of digits of the parameters below is fixed. If the number of digits is not

correct when entering the command, a syntax error results.
• Date

YY/MM/DD (8 characters)

YY: Year (Enter the lower two digits of the year.)

MM: Month

DD: Day

• Time

HH:MM:SS (8 characters)

HH: Hour

MM: Minute

SS: Second

• Channel number: 2 characters (Example: 01, 0A)
• Relay number: 3 characters (Example: I01)
• Communication input data: 3 characters (Example: C02)
• Constants used in the computation function (/M1 option): 3 characters (Example:

K03)

• Remote control (/R1 option) input terminal status: 3 characters (Example: D04)

Chapter 4