3-9

IM 04P01B01-17E

Using

the
RS-422A/485

Communication

Interface

(/C3
Option)

1

2

3

4

5

6

App

Index

3.4 Modbus Slave Specifications

Registers

The registers for using the Modbus slave protocol are listed below. Binary values are

stored to the register in order from the highest byte.

Input Register Data

30001

Measured data of CH01

30002

Measured data of CH02

30003

Measured data of CH03

30004

Measured data of CH04

30005

Measured data of CH05

30006

Measured data of CH06

30007

Measured data of CH07

30008

Measured data of CH08

30009

Measured data of CH09

30010

Measured data of CH10

30011

Measured data of CH11

30012

Measured data of CH12

30013

Measured data of CH13

30014

Measured data of CH14

30015

Measured data of CH15

30016

Measured data of CH16

30017

Measured data of CH17

30018

Measured data of CH18

30019

Measured data of CH19

30020

Measured data of CH20

30021

Measured data of CH21

30022

Measured data of CH22

30023

Measured data of CH23

30024

Measured data of CH24

• The data is a 16-bit signed integer. The value is the same as the measured data in binary

output format (see page 5-14).

• The decimal point and unit information are not included. Set them on the Modbus master.
• Valid channels vary depending on the device. An error response (code 2) occurs when an

invalid channel is read.

Input Register Data

31001

Alarm status of the measured data of CH01

31002

Alarm status of the measured data of CH02

31003

Alarm status of the measured data of CH03

31004

Alarm status of the measured data of CH04

31005

Alarm status of the measured data of CH05

31006

Alarm status of the measured data of CH06

31007

Alarm status of the measured data of CH07

31008

Alarm status of the measured data of CH08

31009

Alarm status of the measured data of CH09

31010

Alarm status of the measured data of CH10

31011

Alarm status of the measured data of CH11

31012

Alarm status of the measured data of CH12

31013

Alarm status of the measured data of CH13

31014

Alarm status of the measured data of CH14

31015

Alarm status of the measured data of CH15

31016

Alarm status of the measured data of CH16

31017

Alarm status of the measured data of CH17

31018

Alarm status of the measured data of CH18

31019

Alarm status of the measured data of CH19

31020

Alarm status of the measured data of CH20

31021

Alarm status of the measured data of CH21

31022

Alarm status of the measured data of CH22

31023

Alarm status of the measured data of CH23

31024

Alarm status of the measured data of CH24

• The value is the same as the alarm status in binary output format (see page 5-14). The data

is entered in the “A2A1A4A3” order in the register.

• Valid channels vary depending on the device. An error response (code 2) occurs when an

invalid channel is read.