App index, Vm sets the message format (/bt1 option) – Yokogawa µR20000 User Manual

Page 62

Advertising
background image

Commands

1

2

3

4

5

6

App

Index

IM 04P01B01-17E

4-19

VP

Turns Start printout/End printout

ON/OFF (/BT1 option).

Syntax

VP p1,p2,p3,p4<terminator>
p1 Mode selection

START Start printou
END

End printout

START2 Start printout 2
END2

End printout 2

p2 Batch name printout ON/OFF (ON, OFF)
p3 Chart speed printout ON/OFF (ON, OFF)
p4 Date/time printout ON/OFF (ON, OFF)

Query

VP[p1]?

Example Turn the Start printout batch name (batch

number and lot number) ON, chart speed OFF,
and date/time ON.
VPSTART,ON,OFF,ON

Description • Valid with the header printout (/BT1 option)

when set to use Start printout/End printout (see
the UE command).

• Start printout 2 and End printout 2 valid when

set to use Start printout 2 and End printout 2
(see the UE command).

VA

Sets the Start printout/End

printout action (/BT1 option).

When setting Start printout

Syntax

VA p1,p2<terminator>
p1 Mode selection

START Start printout
START2 Start printout 2

p2 Feed amount before Start printout (0-50)

[mm]

Query

VA[p1]?

Example Set feed amount before Start printout to 10 mm.

VASTART,10

Description • Valid with the header printout (/BT1 option)

when set to use Start printout/End printout (see
the UE command).

• Start printout 2 valid when set to use Start

printout 2 and End printout 2 (see the UE
command).

When setting End printout

Syntax

VA p1,p2,p3,p4,p5<terminator>
p1 Mode selection

END

End printout

END2

End printout 2

p2 Feed amount after End printout (0-50) [mm]
p3 Lot number automatic updating ON/OFF

(ON, OFF)

p4 Pen offset compensating recording ejection

ON/OFF (ON, OFF)

p5 Select chart speed when ejecting pen offset

compensating record.
C.SPEED Chart speed setting
450mm/h Fixed to 450 mm/h

Query

VA[p1]?

Example Set feed amount after End printout to 10 mm.

VAEND,10

Description • Valid with the header printout (/BT1 option)

when set to use Start printout/End printout (see
the UE command).

• End printout 2 valid when set to use Start

printout 2 and End printout 2 (see the UE
command).

• P3 valid when lot number is set to 4 or 6 digits

(see the UE command).

• P4 valid when POC is set to ON with the pen

model (see the UE command).

• P5 is valid when p4 is ON.

VM

Sets the message format

(/BT1 option).

Syntax

VM p1,p2<terminator>
p1 Message number (1-5)
p2 Message format setting

(µR10000: up to 35 chars., µR20000: up to
69 chars.)
H:M

Hours:minutes (5 characters)

H:M:S Hours:minutes:seconds

(8 characters)

M/D H:M Month/day hours:minutes

(11 characters)

M/D H:M:S Month/day hours:minutes:

seconds (14 characters)

D/M H:M Day/month hours:minutes

(11 characters)

D/M H:M:S Day/month hours:minutes:

seconds (14 characters)

D.M H:M Day.month hours:minutes

(11 characters)

D.M H:M:S Day.month hours:minutes:

seconds (14 characters)

M.D H:M Month.day hours:minutes

(12 characters)

M.D H:M:S Month.day hours:minutes:

seconds (15 characters)

Y/M/D H:M:S Year/month/day hours:

minutes:seconds
(19 characters)

M/D/Y H:M:S Month/day/year:hours:

minutes:seconds
(19 characters)

D/M/Y H:M:S Day/month/year hours:

minutes:seconds
(19 characters)

D.M.Y H:M:S Day.month.year hours:

minutes:seconds
(19 characters)

M.D.Y H:M:S Month/day/year hours:

minutes:seconds
(20 characters)

4.4 Setting Commands

Advertising
This manual is related to the following products: