3-14

IM 04P01B01-17E

3.5 Setting the Serial Interface

The serial interface must be configured.

Setup Items

RS422/485

Address

End

Baud rate

Data length

Parity

Protocol

Procedure

For a description of the basic operations, see “Basic Key Operations” on page vi.

Entering Basic Setting Mode

Hold down the

MENU

key for 3 seconds to display the Setting mode screen.

Next, hold down both the

(

DISP

) key and the

(

FUNC

) key for 3 seconds to

display the Basic Setting mode screen.

Note

To cancel an operation, press the

ESC key.

1.

Press the

key to select

RS422/485 and then press the key.

Basic=RS422/485

2.

Press the

key to select the recorder’s address and then press the key.

Address=1

3.

Press the

key to select the

Baud rate value and then press the key.

Baud rate=38400

4

Press the

key to select the

Data length value and then press the key.

Data length=8

5

Press the

key to select the

Parity value and then press the key.

Parity=Even

6.

Press the

key to select the

NORMAL or MODBUS value and then press the

key.

The setting complete screen appears.

Protocol=NORMAL

RS422/485

Setting complete

Saving the Settings

1.

Press the

ESC key to return to the Basic= screen.

2.

Press the

key to select

End and then press the key.

3.

Press the

key to select

Store and then press the key.

The settings are activated, and the Operation mode screen appears.

* When the key or key is pressed while holding down the

SHIFT key, the operation is reversed as

when the respective key is pressed by itself.