Commands

1

2

3

4

5

6

App

Index

IM 04P01B01-17E

4-11

Description • This command cannot be specified while

computation is in progress.

• Set p4 according to the table in section 4.3.
• For p5 and p6, enter an integer value of 5

digits or less according to the table in section
4.3. The decimal position is fixed to the
position indicated in the table in section 4.3.

• Be sure that p8 is greater than p7.

When setting the square root

Syntax

SR p1,p2,p3,p4,p5,p6,p7,p8,p9,p10
<terminator>

p1 Channel number

p2 Measurement mode (SQRT)

p3 Range type

p4 Leftmost value of span

p5 Rightmost value of span

p6 Leftmost value of scaling (−20000 to 30000)

p7 Rightmost value of scaling (−20000 to

30000)

p8 Scaling decimal point position (0 to 4)

p9 Low-cut function (ON, OFF)

p10 Low-cut value

Query

SR[p1]?

Example Given channel 01 whose input range is 0 to 10 V,

take the square root of the input value and scale
the result in the range 0.00 (m

3

/s) to 100.00

(m

3

/s). When the input value is less than or equal

to 5.0% of the recording span, use the low-cut
function.
SR01,SQRT,20V,0,1000,0,10000,2,ON,50

Description • This command cannot be specified while

computation is in progress.

• Set p3 according to the table in section 4.3.
• For p4 and p5, enter an integer value of 5

digits or less according to the table in section
4.3. The decimal position is fixed to the
position indicated in the table in section 4.3.

• Be sure that p7 is greater than p6.
• Parameters p9 and p10 is valid only when

the low-cut function is enabled (see the UF
command).

VB

Sets the bias.

Syntax

VB p1,p2,p3<terminator>

p1 Channel number

p2 Bias ON/OFF (ON, OFF)

p3 Bias value

Query

VB[p1]?

Example Given channel 03 whose range type is 2V

(measurable range: −2.000 to 2.000 V), add a
bias of 0.1 V.
VB03,ON,100

Description • Set p3 in the range of ±10% of the span of the

measurable range at the range type or ±10%
of the scaling span. Specify the value with an
integer (see section 4.3).

• The bias setting is valid only when the bias

function is enabled (see the UF command).

VL

Sets the calibration correction

(/CC1 option).

Syntax

VL,p1,p2,p3,p4,···, p33,p34
<terminator>

p1 Channel number

p2 Calibration correction function (ON, OFF)

p3 Correction point

p4 Correction value

···

p33 Correction point

p34 Correction value

Query

VL[p1]?

Example Set three sets of correction point and correction

values when channel 02 is set to 2 V range
(measurable range: –2.000 V to 2.000 V) and
the calibration correction mode is set to Revise
Value. (0.000 and 0.001), (1.000 and –0.002),
and (2.000 and 0.001)

VL 02,ON,0,1,1000,-2,2000,1
Set three sets of correction point and correction
values when channel 02 is set to 2 V range
(measurable range: –2.000 V to 2.000 V) and the
calibration correction mode is set to Abs. Value.
VL02,ON,0,1,1000,998,2000,2001

Description • The number of parameters p3 to p34 varies

depending on the number of points specified
by the basic setting command UQ.

• If the input type of the source channel is

VOLT, TC, or RTD, the range of the correction
point and correction values is the same as the
range of the range type (see section 4.3).

• If the measurement mode of the source

channel is SCALE or 1-5V, the range of the
correction point and correction values is –5%
to 105% of the scaling range or –20000 to
30000.
• The following limitations exist.

p3 < p5 ≤ ... ≤ p31 ≤ p33

• During revise value mode

p3+p4 < p5+p6 < ... < p33+p34

• During absolute value mode

p4 < p6 < ... < p32 < p34

SA

Sets the alarm.

When not using the alarm

Syntax

SA p1,p2,p3<terminator>

p1 Channel number

p2 Alarm number (1 to 4)

p3 Alarm ON/OFF state (OFF)

Query

SA[p1[,p2]]?

Example Do not use alarm number 4 of channel 01.

SA01,4,OFF

When using the alarm

Syntax

SA p1,p2,p3,p4,p5,p6,p7<terminator>

p1 Channel number

p2 Alarm number (1 to 4)

p3 Alarm ON/OFF state (ON)

4.4 Setting Commands