Psu series user manual – GW Instek PSU-Series User Manual User Manual
Page 100

PSU Series User Manual
100
V-I Mode
Selects High Speed Priority or Slew Rate
Priority for CV or CC mode. The voltage or
current slew rate can only be edited if CC/CV
Slew Rate Priority is selected. The ISR indicator
will be lit for CC Slew Rate Priority and the
VSR indicator will be lit for CV Slew Rate
Priority.
Note: CC and CV Slew Rate Priority mode are
disabled when voltage/current output is set to
external control.
VOLTAGE
CURRENT
VSR
LAN
RMT
ERR
DLY
ALM
ISR
M 1
M 2
M 3
RUN
C V
C C
V
A
CC Slew Rate priority
VOLTAGE
CURRENT
VSR
LAN
RMT
ERR
DLY
ALM
ISR
M 1
M 2
M 3
RUN
C V
C C
V
A
CV Slew Rate Priority
F-03
0 = CV high speed priority
1 = CC high speed priority
2 = CV slew rate priority
3 = CC slew rate priority
Rising Voltage
Slew Rate
Sets the rising voltage slew rate. Only
applicable if V-I Mode is set to CV Slew Rate
Priority.
F-04
0.001 ~ max. V/msec
Falling Voltage
Slew Rate
Sets the falling voltage slew rate. Only
applicable if V-I Mode is set to CV Slew Rate
Priority.
F-05
0.001 ~ max. V/msec
Rising Current
Slew Rate
Sets the rising current slew rate. Only
applicable if V-I Mode is set to CC Slew Rate
Priority.
F-06
0.001 ~ max. A/msec