GW Instek PSU-Series User Manual User Manual
Page 147

ANALOG CONTROL
147
Connection- alt.
shielding
If the wire shield needs to be grounded at the
voltage source (EXT-V), then it can be
connected as shown below.
PSU
EXT-V
Isolated
Voltage
Option
+
-
4
3
Output
Terminal
2 core shielded
wire or twisted
pair
Pin4(GND_ISO) → EXT-V (-)
Pin3(+IPROG_ISO) → EXT-V (+)
Wire shield → EXT-V ground (GND)
Steps
1. Connect the external voltage according to the
connection diagrams above.
2. Set the F-91 power on
configuration setting to 4 (Control
by Isolated Board).
Page 111
3. Set the F-97 power on
configuration setting to 0 (0~5V
control range) or 1 (0~10V control
range).
Page 111
Be sure to cycle the power after the power
on configuration has been set.
4. Press the Function key and confirm
the new configuration settings (F-
91=4, F-97=0 or 1).
Function