Special function, Setting normal function settings – GW Instek PSU-Series User Manual User Manual
Page 109

CONFIGURATION
109
Special Function
Special Function
The special function setting is used to access
calibration, firmware updates and other special
functions. The special function setting has a
password that is used to access the special
function menu. The password used determines
which function is accessed. Please see your
distributor for details.
F-00
0000 ~ 9999
Setting Normal Function Settings
The Normal Function settings (F-01~F-61, F-
70~F-76) and F-88 ~ F-89 can be easily
configured with the Function key.
Ensure the load is not connected.
Ensure the output is off.
Function settings F-90~98 can only be
viewed.
Note
Function setting F-89 (Show Version) can only be
viewed, not edited.
Configuration settings F-90~ F-98 cannot be edited
in the Normal Function settings. Use the Power On
Configuration settings. See page 111 for details.
Steps
1. Press the Function key. The
function key will light up.
Function
2. The display will show F-01 on the left and the
configuration setting for F-01 on the right.