Output terminals – GW Instek PSU-Series User Manual User Manual
Page 41

OPERATION
41
Output Terminals
Background
Before connecting the output terminals to the
load, first consider whether voltage sense will
be used, the gauge of the cable wiring and the
withstand voltage of the cables and load.
The output terminals can be connected to load
cables using M8 sized bolts.
WARNING
Dangerous voltages. Ensure that the power to the
instrument is disabled before handling the power
supply output terminals. Failing to do so may lead
to electric shock.
Steps
1. Turn the power switch off.
2. Remove the output terminal cover. Page 43
3. If necessary, connect the chassis
ground terminal to either the
positive or negative terminal. See
the grounding chapter for details.
Ground
4. Choose a suitable wire gauge and
crimping terminal for the load
cables.