Voltage sense, Remote sense connector, Age 62 – GW Instek PSU-Series User Manual User Manual
Page 62

PSU Series User Manual
62
Voltage Sense
The PSU power supplies can be operated using local or remote
voltage sense. By default the PSU ships configured for local sense.
Remote Sense Connector
The Remote Sense connector includes a detachable plug to facilitate
making the sense connections. The remote sense connector also has
a safety cover.
WARNING
Ensure the output is off before handling the
remote sense connector.
Use sense cables with a voltage rating exceeding
the isolation voltage of the power supply.
Never connect sensing cables when the output is
on. Electric shock or damage to the power supply
could result.
Remote Sense
Connector
Overview
When using the remote sense connector make
sure the wires that are used follow the following
guidelines:
Wire gauge:
AWG 28 to AWG 16
Strip length
7mm // 0.28 in.
+S
+LS
NC
-LS
-S
+S: Remote(+) sense
+LS: Local (+) sense
NC: Not connected
-LS: Local (-) sense
-S: Remote (-) sense