2. Set the F-90 power on
configuration setting to 4 (Control
by Isolated Board).
Page 111
3. Set the F-97 power on configuration setting to 1.
Be sure to cycle the power after the power
on configuration has been set.
4. Press the Function key and confirm
the new configuration settings (F-
90=4, F-97=1).
Function
5. Press the Output key. The
voltage can now be
controlled with the isolated
external current control.
Output
Note
Use a stable current supply for the external voltage
control.
Note
CV and CC Slew Rate Priority are disabled for V-I
mode (F-03) when using external current control.
See the normal function settings on page 99.
CAUTION
Ensure the polarity is correct when connecting the
external current.
Ensure no more than 21mA are input into the