COMMUNICATION INTERFACE

169

4. Terminate the OUT port of the last unit with

the end terminal connector included in the
PSU-232 or PSU-485 connection kit.

5. Press the Function key to enter the

Normal configuration settings for
the master unit.

Page 109

Set the following settings:

F-70 = 1 or 2

Configure the master unit as you

normally would for RS232 or
RS485 remote control, see page
164.

F-71 = 0~7

Set the baud rate (set all units

the same). See page 164.

F-72 = 1

Set to 8 data bits.

F-73 = 0

Parity to none.

F-74 = 1

1 Stop bit.

F-75 = 1

Set the UART TCP to TDK

(emulation mode).

F-76 = 00~31

Set the address of the master

unit. It must be a unique
address identifier.

6. Press the Function key to enter the

Normal configuration settings for
the slave(s).

Page 109

Set the following settings:

F-70 = 2

Set the slave unit to RS485.

F-71 = 0~7

Set the baud rate (make all

units, including the master, the
same baud). See page 164.

F-72 = 1

Set to 8 data bits.

F-73 = 0

Parity to none.

F-74 = 1

1 Stop bit.

F-75 = 1

Set the UART TCP to TDK

(emulation mode).