Date & time, Password, Clear datalog – Thermo Fisher Scientific Eutech 2700 Series User Manual

Page 12: 8 date & time

Advertising
background image

8

Date & Time

Setting the correct date and time is required for GLP and will apply to power off,

measurement, data log, and print functions. Instrument has battery backup to

retain date/time settings upon power loss. Factory reset will not apply to date and

time setting once it has been set. Changes related to daylight savings time must

be manually entered.

Date Format: Select (MM DD YY) or (DD MM YY)
Time Format: Select 12Hrs (AM/PM) or 24Hrs

Password

Select “

ENABLE” to restrict access to Calibration and Setup modes. When

password protection is enabled, password entry is required before performing any

calibration, or making changes to the setup mode. Setup parameters can be

viewed, but can not be changed without correct password entry. The password is

a user selectable number from 1 to 99999.
Select “

DISABLE” if password protection is not desired.

The meter does not allow you to edit setup parameters or perform a new

calibration unless you enter the correct password. If an incorrect password is

entered 3 consecutive times, the meter returns to measurement mode.
In the event the password can not be recovered, a password can be provided via

a written request to Eutech Instruments/Oakton Instruments. The instrument

serial number and your contact information are required.

Clear Datalog

Select “

YES” to delete all saved data that was manually or automatically stored

into memory. This step is most commonly required when the “TIMED” data log

interval is not turned off and as a result, fills the memory to capacity. When the

memory is full, deleting the saved data is required in order to save additional

data/ The 2700 will never erase old data in favor of new data when the memory is

full.
If you wish to view, print or send the saved data to a PC before deleting, use the

MEM OUT” function from measurement mode.