Cal points, Resolution, Alarm – Thermo Fisher Scientific Eutech 2700 Series User Manual

Page 15: Cal due, 11 cal points

Advertising
background image

11

Cal Points

Specify the number of pH calibration points you intend to

calibrate with. Select the number of calibration points

from 1 to 5 with preset buffer groups or from 2 to 5

points with custom buffer group.
For example, if you have selected the USA buffer group

which contains 5 standards, but you will only be using 3

standards (pH 4, 7, and 10), selecting 3 calibration

points here will speed up the calibration process—the 2700 will automatically

recognize calibration is finished after the 3

rd

point has been completed.

Resolution

Select your desired pH resolution;

0.0, 0.00, or 0.000.

Tip: Don’t choose more than you

really need—0.00 is the factory default setting since it is

most commonly used. Choosing 0.000 will lead to more

“unstable” conditions and cause a delay in obtaining the

final reading as the last digit (which is often not needed)

changes.

Alarm

The 2700 offers a visual and audible alarm to alert you

when the High or Low values that have been set from

this menu are exceeded.

“High Alarm” or “Low Alarm” will blink on the display if

the values are exceeded while simultaneously, a loud,

intermittent beeping sound is heard. The alarms will continue until the conditions

are no longer met, and will only be active during measurement mode.

Cal Due

When enabled, the “Cal Due” indicator blinks if the

number of days since the last calibration has been

exceeded. Set the number of days from 1 to 31.

Advertising