Adjustment) – Thermo Fisher Scientific Eutech 2700 Series User Manual

Page 29

Advertising
background image

25

12.

Conductivity, TDS, Salinity, & Resistivity

Calibration (manual adjustment)

For best results, periodic calibration with known accurate standards is

recommended prior to measurement. Calibrate with standards that are close in

value to your intended sample(s).
When using multi-point calibration, perform a calibration in each range that you

expect to use for best results.
If the electrode has been stored dry for some time, soaking in alcohol or clean

water for a few minutes can help performance.
1) Turn meter on, press MODE if needed for the appropriate mode.

2) Press CAL if needed to enter calibration mode. Rinse the electrode with

clean water then submerse into your calibration standard.

3) Use the up/down arrows to adjust the cell constant manually or press NEXT

to keep the nominal cell constant value and proceed with a typical manual

calibration (recommended). If adjusting the cell constant manually, press

ENTER after ‘Stable’ appears to confirm the adjustment.

Before and after a manual calibration constant adjustment

4) The upper display is the active reading while the lower display is the factory

default value without calibration. Use the up/down arrows to adjust the upper

display to match the desired calibration value. Press

ENTER after ‘Stable’

appears to confirm the calibration. The calibrated value, Range (R) that was

calibrated and the calculated cell constant are now shown.